Richter P H, Steinbrener J, Schicho A, Gebhard F
Department for Orthopaedic Trauma, Hand and Reconstructive Surgery, Ulm University, Albert-Einstein-Allee 23, D-89081 Ulm, Germany.
Siemens Healthcare GmbH, Henkestr. 172, 91052 Erlangen, Germany.
Injury. 2016 Aug;47(8):1608-12. doi: 10.1016/j.injury.2016.04.031. Epub 2016 Apr 23.
Mobile C-arm imaging is commonly used in operating rooms worldwide. Especially in orthopaedic surgery, intraoperative C-arms are used on a daily basis. Because of new minimally-invasive surgical procedures a development in intraoperative imaging is required. The purpose of this article is investigate if the choice of mobile C-arms with flat panel detector technology (Siemens Cios Alpha and Ziehm Vision RFD) influences image quality and dose using standard, commercially available test devices.
For a total of four clinical application settings, two zoom formats, and all dose levels provided, the transmission dose was measured and representative images were recorded for each test device. The data was scored by four observers to assess low contrast and spatial resolution performance. The results were converted to a relative image quality figure allowing for a direct image quality and dose comparison of the two systems.
For one test device, the Cios Alpha system achieved equivalent (within the inter-observer standard error) or better low contrast resolution scores at significantly lower dose levels, while the results of the other test device suggested that both systems achieved similar image quality at the same dose. The Cios Alpha system achieved equivalent or better spatial resolution at significantly lower dose for all application settings except for Cardiac, where a comparable spatial resolution was achieved at the same dose.
The correct choice of a mobile C-arm is very important, because it can lead to a reduction of the intraoperative radiation dose without negative effects on image quality. This can be a big advantage to reduce intraoperative radiation not only for the patient but also for the entire OR-team.
移动C形臂成像在全球手术室中普遍使用。特别是在骨科手术中,术中每天都会使用C形臂。由于新的微创手术程序,需要对术中成像进行改进。本文的目的是使用标准的市售测试设备,研究采用平板探测器技术的移动C形臂(西门子Cios Alpha和齐默Vision RFD)的选择是否会影响图像质量和剂量。
对于总共四种临床应用设置、两种变焦格式以及所提供的所有剂量水平,测量传输剂量,并为每个测试设备记录代表性图像。由四名观察者对数据进行评分,以评估低对比度和空间分辨率性能。结果被转换为相对图像质量指标,以便对两个系统的图像质量和剂量进行直接比较。
对于一个测试设备,Cios Alpha系统在显著更低的剂量水平下实现了等效(在观察者间标准误差范围内)或更好的低对比度分辨率分数,而另一个测试设备的结果表明,两个系统在相同剂量下实现了相似的图像质量。除心脏应用外,Cios Alpha系统在所有应用设置下均在显著更低的剂量下实现了等效或更好的空间分辨率,在心脏应用中,在相同剂量下实现了可比的空间分辨率。
正确选择移动C形臂非常重要,因为它可以在不影响图像质量的情况下降低术中辐射剂量。这对于减少术中辐射不仅对患者而且对整个手术室团队来说可能是一个很大的优势。