三种寰枢椎主动预防椎动脉损伤固定组合的生物力学研究

Biomechanical Study of 3 Atlantoaxial Proactive Vertebral Artery Injury Prevention Fixation Combinations.

作者信息

Zhang Peng, Liu Qi, Xu Tianming, Sun Dezhi, Yang Jun, Ni Bin

机构信息

Department of Orthopedics, The Second Affiliated Hospital, The Second Military Medical University, Shanghai, China.

Department of Orthopedics, The First Affiliated Hospital of Xiamen University, Xiamen, China.

出版信息

World Neurosurg. 2016 Sep;93:293-8. doi: 10.1016/j.wneu.2016.06.061. Epub 2016 Jun 23.

Abstract

OBJECTIVE

To evaluate the biomechanical stability of 3 atlantoaxial proactive vertebral artery injury prevention fixation combinations.

METHODS

We used 6 fresh human cadaveric C0-C4 specimens for the biomechanical test with moment control in 6 loading directions. First, range of motion (ROM) was measured in the intact specimens before they were destabilized; next, ROM of the 4 conditions fixed with 4 different fixation combinations were tested in alternating sequence: bilateral transarticular screw (TAS) and 3 test combinations: C1 laminar hook (C1LH) and C2 intralaminar screw (C2ILS) on 1 side combined with TAS (C1LH-C2ILS/TAS) or C1LH and TAS (C1LH-C2ILS/C1LH-TAS) or C1LH and C2ILS on the other side (C1LH-C2ILS/C1LH-C2ILS).

RESULTS

ROM of all 4 instrumented sets significantly decreased compared with intact specimens in all the loading cases (flexion, left and right lateral bending, left and right axial rotation) except extension. Compared with the TAS group, all 3 test groups showed higher stiffness in flexion and equivalent stability in the other 5 directions. Side-by-side comparison among the 3 test fixation techniques showed no significant difference.

CONCLUSIONS

The combination of C1LH and C2ILS supplemented with contralateral TAS or C1LH and TAS or C1LH and C2ILS was superior to bilateral TAS fixation with regard to biomechanics and vertebral artery safety.

摘要

目的

评估三种寰枢椎主动预防椎动脉损伤固定组合的生物力学稳定性。

方法

我们使用6个新鲜的人体C0 - C4标本,在6个加载方向上进行力矩控制的生物力学测试。首先,在标本失稳前测量完整标本的活动范围(ROM);接下来,对用4种不同固定组合固定的4种情况的ROM进行交替测试:双侧经关节螺钉(TAS)和3种测试组合:一侧的C1椎板钩(C1LH)和C2椎板内螺钉(C2ILS)与TAS联合(C1LH - C2ILS/TAS)或C1LH和TAS(C1LH - C2ILS/C1LH - TAS)或另一侧的C1LH和C2ILS(C1LH - C2ILS/C1LH - C2ILS)。

结果

在所有加载情况下(屈曲、左右侧屈、左右轴向旋转),除伸展外,所有4种器械固定组的ROM与完整标本相比均显著降低。与TAS组相比,所有3个测试组在屈曲时表现出更高的刚度,在其他5个方向上稳定性相当。3种测试固定技术的并排比较显示无显著差异。

结论

在生物力学和椎动脉安全性方面,C1LH和C2ILS联合对侧TAS或C1LH和TAS或C1LH和C2ILS优于双侧TAS固定。