Yoshida M, Morimoto S, Uekado Y, Yasukawa S, Aoshi H, Yoshida T, Ebisuno S, Ohkawa T
Department of Urology, Wakayama Medical College.
Hinyokika Kiyo. 1989 Feb;35(2):217-23.
The reliability of measurement of urinary TPA using a Prolifigen TPA kit (Sangtec, Sweden) was investigated, and an improved method for stabilizing TPA was tested. Samples were filtrated with a cellulose nitrate membrane filter to remove impurities as a possible cause of a high false positive rate which has been a problem in urinary TPA measurement. Serum TPA and Tween 40-treated urinary TPA was not adsorbed by the filter, whereas TPA in the non-treated urine showed a marked affinity to the filter. This shows the instability of TPA in the original urine, and the impropriety of filtration as a measure of purification of urine samples. The dilution test of urine and the standard TPA, and the recovery test of the standard TPA added to 6 different urine samples showed reasonable results only when the samples were treated with Tween 40; otherwise the values were lower than expected. Thirty-four spot urine samples were tested to examine the change in TPA value during the storage for 24 hours. TPA values decreased when urine was stored at room temperature, but this change was prevented by treatment of urine with Tween 40.
研究了使用Prolifigen TPA试剂盒(瑞典Sangtec公司)测量尿TPA的可靠性,并测试了一种稳定TPA的改进方法。用硝酸纤维素膜过滤器过滤样品以去除杂质,杂质可能是尿TPA测量中高假阳性率这一问题的原因。血清TPA和吐温40处理的尿TPA不被过滤器吸附,而未处理尿液中的TPA对过滤器显示出明显的亲和力。这表明原始尿液中TPA的不稳定性,以及过滤作为尿液样品纯化措施的不当性。尿液和标准TPA的稀释试验,以及添加到6个不同尿液样品中的标准TPA的回收率试验仅在样品用吐温40处理时显示出合理的结果;否则,值低于预期。测试了34份随机尿样,以检查储存24小时期间TPA值的变化。当尿液在室温下储存时,TPA值下降,但通过用吐温40处理尿液可防止这种变化。