脉冲序列参数选择对基于MRI的前列腺植入后评估中阳性对比MRI标记物信号强度的影响。

Effect of pulse sequence parameter selection on signal strength in positive-contrast MRI markers for MRI-based prostate postimplant assessment.

作者信息

Lim Tze Yee, Kudchadker Rajat J, Wang Jihong, Stafford R Jason, MacLellan Christopher, Rao Arvind, Ibbott Geoffrey S, Frank Steven J

机构信息

Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030 and The University of Texas at Houston Graduate School of Biomedical Sciences, 6767 Bertner Avenue, Houston, Texas 77030.

Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030.

出版信息

Med Phys. 2016 Jul;43(7):4312. doi: 10.1118/1.4953635.

Abstract

PURPOSE

For postimplant dosimetric assessment, computed tomography (CT) is commonly used to identify prostate brachytherapy seeds, at the expense of accurate anatomical contouring. Magnetic resonance imaging (MRI) is superior to CT for anatomical delineation, but identification of the negative-contrast seeds is challenging. Positive-contrast MRI markers were proposed to replace spacers to assist seed localization on MRI images. Visualization of these markers under varying scan parameters was investigated.

METHODS

To simulate a clinical scenario, a prostate phantom was implanted with 66 markers and 86 seeds, and imaged on a 3.0T MRI scanner using a 3D fast radiofrequency-spoiled gradient recalled echo acquisition with various combinations of scan parameters. Scan parameters, including flip angle, number of excitations, bandwidth, field-of-view, slice thickness, and encoding steps were systematically varied to study their effects on signal, noise, scan time, image resolution, and artifacts.

RESULTS

The effects of pulse sequence parameter selection on the marker signal strength and image noise were characterized. The authors also examined the tradeoff between signal-to-noise ratio, scan time, and image artifacts, such as the wraparound artifact, susceptibility artifact, chemical shift artifact, and partial volume averaging artifact. Given reasonable scan time and managable artifacts, the authors recommended scan parameter combinations that can provide robust visualization of the MRI markers.

CONCLUSIONS

The recommended MRI pulse sequence protocol allows for consistent visualization of the markers to assist seed localization, potentially enabling MRI-only prostate postimplant dosimetry.

摘要

目的

在植入后剂量学评估中,计算机断层扫描(CT)常用于识别前列腺近距离放射治疗种子源,但代价是牺牲了精确的解剖轮廓勾画。磁共振成像(MRI)在解剖学描绘方面优于CT,但识别阴性对比种子源具有挑战性。有人提出使用阳性对比MRI标记物取代间隔物,以协助在MRI图像上定位种子源。研究了在不同扫描参数下这些标记物的可视化情况。

方法

为模拟临床场景,在前列腺模型中植入66个标记物和86个种子源,并在3.0T MRI扫描仪上成像,使用三维快速射频扰相梯度回波采集,采用各种扫描参数组合。系统地改变扫描参数,包括翻转角、激励次数、带宽、视野、层厚和编码步数,以研究它们对信号、噪声、扫描时间、图像分辨率和伪影的影响。

结果

表征了脉冲序列参数选择对标记物信号强度和图像噪声的影响。作者还研究了信噪比、扫描时间和图像伪影(如卷绕伪影、磁化率伪影、化学位移伪影和部分容积平均伪影)之间的权衡。在合理的扫描时间和可管理的伪影条件下,作者推荐了能够提供MRI标记物可靠可视化的扫描参数组合。

结论

推荐的MRI脉冲序列方案能够一致地可视化标记物以协助种子源定位,有可能实现仅使用MRI的前列腺植入后剂量学。

相似文献

引用本文的文献

本文引用的文献

5
Localization of brachytherapy seeds in MRI by deconvolution.通过反卷积在磁共振成像中定位近距离治疗种子源。
Annu Int Conf IEEE Eng Med Biol Soc. 2013;2013:2960-3. doi: 10.1109/EMBC.2013.6610161.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验