多色流式细胞术在检测小的单克隆浆细胞群体方面比显微镜检查更敏感。

Polychromatic flow cytometry is more sensitive than microscopy in detecting small monoclonal plasma cell populations.

作者信息

Tran Daniel N, Smith Sandy A B C, Brown David A, Parker Andrew J C, Joseph Joanne E, Armstrong Nicola, Sewell William A

机构信息

Immunology Department, SydPath, St Vincent's Hospital Sydney, New South Wales, Australia.

St Vincent's Clinical School, University of New South Wales, New South Wales, Australia.

出版信息

Cytometry B Clin Cytom. 2017 Mar;92(2):136-144. doi: 10.1002/cyto.b.21401. Epub 2016 Aug 5.

Abstract

BACKGROUND

There is an emerging role for flow cytometry (FC) in the assessment of small populations of plasma cells (PC). However, FC's utility has been questioned due to consistent underestimation of the percentage of PC compared to microscopy.

METHODS

A retrospective study was performed on bone marrow samples analysed by 8-colour FC. Plasma cell populations were classified as polyclonal or monoclonal based on FC analysis. FC findings were compared with microscopy of aspirates, histology and immunohistochemistry of trephine biopsies, and immunofixation (IFX) of serum and/or urine.

RESULTS

FC underestimated PC compared to aspirate and trephine microscopy. The 10% diagnostic cutoff for MM on aspirate microscopy corresponded to a 3.5% cutoff on FC. Abnormal plasma cell morphology by aspirate microscopy and clonality by FC correlated in 229 of 294 cases (78%). However, in 50 cases, FC demonstrated a monoclonal population but microscopy reported no abnormality. In 15 cases, abnormalities were reported by microscopy but not by FC. Clonality assessment by trephine microscopy and FC agreed in 251/280 cases (90%), but all 29 discordant cases were monoclonal by FC and not monoclonal by microscopy. These cases had fewer PC and proportionally more polyclonal PC, and when IFX detected a paraprotein, it had the same light chain as in the PC determined by FC.

CONCLUSIONS

FC was more sensitive in detecting monoclonal populations that were small or accompanied by polyclonal PC. This study supports the inclusion of FC in the evaluation of PC, especially in the assessment of small populations. © 2016 International Clinical Cytometry Society.

摘要

背景

流式细胞术(FC)在评估少量浆细胞(PC)群体方面正发挥着越来越重要的作用。然而,由于与显微镜检查相比,FC对PC百分比的估计一直偏低,其效用受到了质疑。

方法

对通过8色FC分析的骨髓样本进行回顾性研究。根据FC分析将浆细胞群体分类为多克隆或单克隆。将FC结果与抽吸物显微镜检查、环钻活检的组织学和免疫组织化学以及血清和/或尿液的免疫固定(IFX)结果进行比较。

结果

与抽吸物和环钻显微镜检查相比,FC低估了PC。抽吸物显微镜检查中MM的10%诊断临界值对应于FC的3.5%临界值。抽吸物显微镜检查显示的异常浆细胞形态与FC显示的克隆性在294例中的229例(78%)中相关。然而,在50例中,FC显示为单克隆群体,但显微镜检查报告无异常。在15例中,显微镜检查报告有异常,但FC未报告。环钻显微镜检查和FC的克隆性评估在280例中的251例(90%)中一致,但所有29例不一致的病例在FC下为单克隆,在显微镜下不是单克隆。这些病例中的PC较少,多克隆PC比例较高,当IFX检测到副蛋白时,其轻链与FC确定的PC中的轻链相同。

结论

FC在检测小的或伴有多克隆PC的单克隆群体方面更敏感。本研究支持将FC纳入PC的评估中,尤其是在评估少量群体时。©2016国际临床细胞计量学会。