Quality requirements for veterinary hematology analyzers in small animals-a survey about veterinary experts' requirements and objective evaluation of analyzer performance based on a meta-analysis of method validation studies: bench top hematology analyzer.

作者信息

Cook Andrea M, Moritz Andreas, Freeman Kathleen P, Bauer Natali

机构信息

Department of Clinical Pathophysiology and Clinical Pathology, Faculty of Veterinary Medicine, Justus-Liebig-University, Gießen, Germany.

IDEXX Laboratories, Ltd, Grange House, Wetherby, UK.

出版信息

Vet Clin Pathol. 2016 Sep;45(3):466-76. doi: 10.1111/vcp.12383. Epub 2016 Jul 18.

Abstract

BACKGROUND

Scarce information exists about quality requirements and objective evaluation of performance of large veterinary bench top hematology analyzers.

OBJECTIVE

The study was aimed at comparing the observed total error (TEobs ) derived from meta-analysis of published method validation data to the total allowable error (TEa ) for veterinary hematology variables in small animals based on experts' opinions. Ideally, TEobs should be < TEa .

METHODS

An online survey was sent to veterinary experts in clinical pathology and small animal internal medicine for providing the maximal allowable deviation from a given result for each variable. Percent of TEa = (allowable median deviation/clinical threshold) * 100%. Second, TEobs for 3 laser-based bench top hematology analyzers (ADVIA 2120; Sysmex XT2000iV, and CellDyn 3500) was calculated based on method validation studies published between 2005 and 2013 (n = 4). The percent TEobs = 2 * CV (%) + bias (%). The CV was derived from published studies except for the ADVIA 2120 (internal data), and bias was estimated from the regression equation.

RESULTS

A total of 41 veterinary experts (19 diplomates, 8 residents, 10 postgraduate students, 4 anonymous specialists) responded. The proposed range of TEa was wide, but generally ≤ 20%. The TEobs was < TEa for all variables and analyzers except for canine and feline HGB (high bias, low CV) and platelet counts (high bias, high CV).

CONCLUSIONS

Overall, veterinary bench top analyzers fulfilled experts' requirements except for HGB due to method-related bias, and platelet counts due to known preanalytic/analytic issues.

摘要

背景

关于大型兽医台式血液分析仪的质量要求和性能客观评估的信息稀缺。

目的

本研究旨在将已发表方法验证数据的荟萃分析得出的观察总误差(TEobs)与基于专家意见的小动物兽医血液学变量的总允许误差(TEa)进行比较。理想情况下,TEobs应小于TEa。

方法

向临床病理学和小动物内科学领域的兽医专家发送了一项在线调查,以获取每个变量与给定结果的最大允许偏差。TEa百分比=(允许的中位数偏差/临床阈值)×100%。其次,根据2005年至2013年发表的方法验证研究(n = 4)计算了3种基于激光的台式血液分析仪(ADVIA 2120;Sysmex XT2000iV和CellDyn 3500)的TEobs。TEobs百分比=2×CV(%)+偏差(%)。除ADVIA 2120(内部数据)外,CV来自已发表的研究,偏差由回归方程估算。

结果

共有41位兽医专家(19位专科医师、8位住院医师、10位研究生、4位匿名专家)做出回应。提议的TEa范围很广,但一般≤20%。除犬猫血红蛋白(高偏差、低CV)和血小板计数(高偏差、高CV)外,所有变量和分析仪的TEobs均小于TEa。

结论

总体而言,兽医台式分析仪除因方法相关偏差导致的血红蛋白以及因已知的分析前/分析问题导致的血小板计数外,均符合专家要求。