Sentkowska Aleksandra, Biesaga Magdalena, Pyrzynska Krystyna
Department of Chemistry, University of Warsaw, Pasteura 1, 02-093 Warsaw, Poland.
J Anal Methods Chem. 2016;2016:3789348. doi: 10.1155/2016/3789348. Epub 2016 Jun 28.
Hydrophilic interaction chromatography (HILIC) was employed to investigate chromatographic behavior of selected flavonoids from their different subgroups differing in polarity. Chromatographic measurements were performed on two different HILIC columns: unmodified silica (Atlantis-HILIC) and zwitterionic sulfoalkylbetaine (SeQuant ZIC-HILIC). Separation parameters such as content and type of organic modifier were studied. On ZIC column retention factors were observed to be inversely proportional to the buffer content in the mobile phase, which is the typical partitioning mechanism. In the case of bare silica column more or less apparent dual retention mechanism was observed, depending on the water component content in the mobile phase. ZIC-HILIC showed better selectivity (in comparison to silica column) with the detection limit of 0.01 mg/L (only for rutin was 0.05 mg/L). Finally, this chromatographic procedure was validated and applied for the determination of some flavonoids in Genista tinctoria L. extract.
采用亲水作用色谱法(HILIC)研究了不同亚组中极性不同的选定黄酮类化合物的色谱行为。在两种不同的HILIC柱上进行色谱测定:未改性硅胶柱(Atlantis-HILIC)和两性离子磺基甜菜碱柱(SeQuant ZIC-HILIC)。研究了诸如有机改性剂的含量和类型等分离参数。在ZIC柱上,保留因子与流动相中的缓冲液含量成反比,这是典型的分配机制。在裸硅胶柱的情况下,根据流动相中的水成分含量,观察到或多或少明显的双重保留机制。与硅胶柱相比,ZIC-HILIC表现出更好的选择性,检测限为0.01 mg/L(仅芦丁为0.05 mg/L)。最后,对该色谱方法进行了验证,并应用于染料木提取物中某些黄酮类化合物的测定。