Busa Lori Shayne Alamo, Komatsu Takeshi, Mohammadi Saeed, Maeki Masatoshi, Ishida Akihiko, Tani Hirofumi, Tokeshi Manabu
Graduate School of Chemical Sciences and Engineering, Hokkaido University.
Anal Sci. 2016;32(8):815-8. doi: 10.2116/analsci.32.815.
We report on the colorimetric oxidation of 3,3',5,5'-tetramethylbenzidine (TMB) by hydrogen peroxide using horseradish peroxidase on photolithography-fabricated (P-PAD) and wax-printed (W-PAD) paper-based analytical devices. Fabricating PADs via photolithography exposes the hydrophilic areas to polymers (photoresists) and solvents, not only reducing the hydrophilicity, but also affecting the TMB-H2O2 assay system with an unavoidable incomplete elimination of photoresist during fabrication. Detection signals are then observed in the presence of photoresist residues on the P-PAD, even at a blank HRP concentration.