Stress shielding is a well-recognized complication associated with total knee arthroplasty. However, this phenomenon has not been thoroughly described. Specifically, no study to our knowledge has evaluated the radiographic impact of utilizing various tibial component compositions on tibial stress shielding.
We retrospectively reviewed 3 cohorts of 50 patients that had a preoperative varus deformity and were implanted with a titanium, cobalt chromium (CoCr), or an all polyethylene tibial implant. A radiographic comparative analysis was performed to evaluate the amount of medial tibial bone loss in each cohort. In addition, a clinical outcomes analysis was performed on the 3 cohorts.
The CoCr was noted to have a statistically significant increase in medial tibial bone loss compared with the other 2 cohorts. The all polyethylene cohort had a statistically significantly higher final Knee Society Score and was associated with the least amount of stress shielding.
The CoCr tray is the most rigid of 3 implants that were compared in this study. Interestingly, this cohort had the highest amount of medial tibial bone loss. In addition, 1 patient in the CoCr cohort had medial soft tissue irritation which was attributed to a prominent medial tibial tray which required revision surgery to mitigate the symptoms.
应力遮挡是全膝关节置换术公认的一种并发症。然而,这一现象尚未得到充分描述。具体而言,据我们所知,尚无研究评估使用不同胫骨假体成分对胫骨应力遮挡的影像学影响。
我们回顾性分析了3组各50例术前有内翻畸形且植入钛合金、钴铬合金(CoCr)或全聚乙烯胫骨假体的患者。进行影像学对比分析,以评估每组内侧胫骨骨量丢失情况。此外,对这3组患者进行了临床疗效分析。
与其他两组相比,CoCr组内侧胫骨骨量丢失有统计学意义的显著增加。全聚乙烯组最终膝关节协会评分在统计学上显著更高,且应力遮挡量最少。
在本研究对比的3种植入物中,CoCr托盘最硬。有趣的是,该组内侧胫骨骨量丢失最多。此外,CoCr组有1例患者出现内侧软组织刺激,原因是内侧胫骨托盘突出,需要翻修手术以缓解症状。