Park Jung Wook, Cheon Min Woo, Lee Min Hong
Topteam Rehabilitation Clinic, Gwangju, Korea.
Department of Hospital Biomedical Engineering, Dongshin University, Gwangju, Korea.
Ann Rehabil Med. 2016 Aug;40(4):575-82. doi: 10.5535/arm.2016.40.4.575. Epub 2016 Aug 24.
To compare the visibility and procedural parameters between a standard spinal needle and a new laser-etched needle (LEN) in real-time ultrasonography guided lumbar medial branch access in a phantom of the lumbosacral spine.
We conducted a prospective single-blinded observational study at a rehabilitation medicine center. A new model of LEN was manufactured with a standard 22-gauge spinal needle and a laser etching machine. Thirty-two inexperienced polyclinic medical students performed ultrasonography-guided lumbar medial branch access using both a standard spinal needle and a LEN with scanning protocol. The outcomes included needle visibility score, needle elapsed time, first-pass success rate, and number of needle sticks.
The LEN received significantly better visibility scores and shorter needle elapsed time compared to the standard spinal needle. First-pass success rate and the number of needle sticks were not significantly different between needles.
A new LEN is expected to offer better visibility and enable inexperienced users to perform an ultrasonography-guided lumbar medial branch block more quickly. However, further study of variables may be necessary for clinical application.
在腰骶椎模型中,比较标准脊髓穿刺针与新型激光蚀刻针(LEN)在实时超声引导下腰椎内侧支穿刺时的可视性和操作参数。
我们在一家康复医学中心进行了一项前瞻性单盲观察研究。使用标准的22号脊髓穿刺针和激光蚀刻机制造了一种新型的LEN。32名缺乏经验的门诊医学生按照扫描方案,使用标准脊髓穿刺针和LEN进行超声引导下腰椎内侧支穿刺。结果包括针的可视性评分、进针时间、首次穿刺成功率和穿刺次数。
与标准脊髓穿刺针相比,LEN的可视性评分明显更高,进针时间更短。两种针的首次穿刺成功率和穿刺次数无显著差异。
新型LEN有望提供更好的可视性,并使缺乏经验的使用者能够更快地进行超声引导下腰椎内侧支阻滞。然而,临床应用可能需要对变量进行进一步研究。