Pazzaglia U E, Pedrotti L, Ramella R, Zatti G
Istituto di Clinica Ortopedica e Traumatologica dell'Università, I.R.C.C.S. Policlinico S. Matteo, Pavia.
Ital J Orthop Traumatol. 1989 Jun;15(2):221-9.
A study of 60 femoral prosthetic components removed after 1-20 years allowed us to study the behaviour of the metallic surface in real life conditions. Changes were observed exclusively in cemented prostheses. They consisted of corrosion phenomena characteristic of the components of steel EN 58 J and 316 S, and wear phenomena secondary to mobilisation. The latter were observed in steel components as well as in those in Co-Cr alloy. The areas of corrosion in the steel prostheses were characteristically distributed, indicating a combination of causal factors, including an insufficient capacity of passivation of the metal and the particular situation of the cement-to-metal interface.