Fracture of totally implanted central venous access devices: a propensity-score-matched comparison of risks for Groshong silicone versus polyurethane catheters.

作者信息

Kojima Soichi, Hiraki Takao, Gobara Hideo, Iguchi Toshihiro, Fujiwara Hiroyasu, Matsui Yusuke, Mitsuhashi Toshiharu, Kanazawa Susumu

机构信息

Department of Radiology, Okayama University Medical School, Okayama - Japan.

Department of Epidemiology, Okayama University Medical School, Okayama - Japan.

出版信息

J Vasc Access. 2016 Nov 2;17(6):535-541. doi: 10.5301/jva.5000606. Epub 2016 Oct 21.

Abstract

PURPOSE

To evaluate retrospectively the fracture risk of totally implanted venous access devices connected to Groshong silicone (SC) versus polyurethane (PU) catheters, inserted via the internal jugular vein.

MATERIALS AND METHODS

The study population comprised 384 SC and 221 PU central venous catheters implanted via the internal jugular vein. The presence of catheter fracture was evaluated. Variables possibly related to catheter fracture were evaluated. First, in order to determine the factors associated with fracture, fracture rates were compared with the log-rank test between the two groups divided by each of the variables. Then, in order to adjust for potential confounders, propensity-score matching of the variables was employed in the two catheter groups. Finally, the rates of fracture were compared between the two propensity-score-matched catheter groups.

RESULTS

There were 16 cases of catheter fracture, for an overall fracture percentage of 2.6% (16/605). All 16 cases of fracture occurred in the SC catheter group. Smaller patient body mass index (p = 0.039), deeper catheter tip position (p = 0.022), and SC catheters (p = 0.019) were significantly associated with fracture. With the propensity-score-matching method, 180 cases were selected in each catheter group. Comparison of the two propensity-score-matched groups showed that fracture rates for SC catheters remained significantly (p = 0.018) higher than those for PU catheters.

CONCLUSIONS

Ports connected to Groshong SC catheters - when implanted via the internal jugular vein - posed a higher risk of fracture than did ports connected to PU catheters.

摘要

目的

回顾性评估经颈内静脉插入的连接Groshong硅胶(SC)导管与聚氨酯(PU)导管的完全植入式静脉通路装置的断裂风险。

材料与方法

研究人群包括经颈内静脉植入的384根SC中心静脉导管和221根PU中心静脉导管。评估导管断裂情况。评估可能与导管断裂相关的变量。首先,为确定与断裂相关的因素,通过对数秩检验比较按每个变量划分的两组之间的断裂率。然后,为调整潜在混杂因素,在两组导管中对变量进行倾向得分匹配。最后,比较两个倾向得分匹配的导管组之间的断裂率。

结果

有16例导管断裂,总体断裂百分比为2.6%(16/605)。所有16例断裂均发生在SC导管组。较小的患者体重指数(p = 0.039)、较深的导管尖端位置(p = 0.022)和SC导管(p = 0.019)与断裂显著相关。采用倾向得分匹配法,每组导管各选取180例。两个倾向得分匹配组的比较显示,SC导管的断裂率仍显著高于PU导管(p = 0.018)。

结论

经颈内静脉植入时,连接Groshong SC导管的端口比连接PU导管的端口具有更高的断裂风险。