Suppr超能文献

320层容积CT扫描仪上的腔内衰减梯度冠状动脉CT血管造影:使用造影剂流模体研究扫描时间、冠状动脉狭窄和心输出量的影响

Transluminal attenuation-gradient coronary CT angiography on a 320-MDCT volume scanner: Effect of scan timing, coronary artery stenosis, and cardiac output using a contrast medium flow phantom.

作者信息

Funama Yoshinori, Utsunomiya Daisuke, Oda Seitaro, Shimonobo Toshiaki, Nakaura Takeshi, Mukunoki Toshifumi, Kidoh Masafumi, Yuki Hideaki, Yamashita Yasuyuki

机构信息

Department of Medical Physics, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan.

Department of Diagnostic Radiology, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan.

出版信息

Phys Med. 2016 Nov;32(11):1415-1421. doi: 10.1016/j.ejmp.2016.10.011. Epub 2016 Oct 22.

Abstract

PURPOSE

Transluminal-attenuation-gradient (TAG) may reflect patient characteristics and physiological parameters. Furthermore, TAG may be affected by factors such as the CT scanner speed, scanning method, scan timing after contrast-medium (CM) injection, and the injection methods. The purpose of our study was to investigate quantitative TAG at different scan timing points after CM injection for coronary CT angiography.

MATERIALS AND METHODS

Using a CM flow phantom and two types of connecting tube mimicking 0% and 70% coronary artery stenosis, we performed 320-detector volume scanning. The heart rate was set at 60bpm and cardiac-output (CO) at 2.0 and 4.0 l/min, respectively. The acquisition time repeated at 0.5-s intervals for 40 and 25s at a CO of 2.0- and 4.0 l/min. We measured the CT number on the same slice level, calculated the time-density-curve (TDC) and the TAG at each time point.

RESULTS

At COs of 2.0 and 4.0 l/min at 0% stenosis, TAG exhibited smaller variations (-3.02 to +0.55HU/cm at 2.0 l/min, -2.63 to +0.43HU/cm at 4.0 l/min) than at 70% stenosis at each time point along the TDC. Compared with a CO at 2.0 l/min with 70% stenosis, the TAG curve for a CO at 4.0 l/min gradually changed with time (-6.64 to +1.18HU/cm at 2.0 l/min vs. -3.46 to +2.75HU/cm at 4.0 l/min).

CONCLUSION

The TAG value was affected by scan timing after CM injection and by CO although the size of the connecting tube with and without stenosis was identical.

摘要

目的

腔内衰减梯度(TAG)可能反映患者特征和生理参数。此外,TAG可能受CT扫描仪速度、扫描方法、造影剂(CM)注射后的扫描时间以及注射方法等因素影响。我们研究的目的是在冠状动脉CT血管造影中,研究CM注射后不同扫描时间点的定量TAG。

材料与方法

使用CM流动模型和两种模拟0%和70%冠状动脉狭窄的连接管,进行320排容积扫描。心率设定为60次/分钟,心输出量(CO)分别设定为2.0和4.0升/分钟。在CO为2.0升/分钟和4.0升/分钟时,采集时间以0.5秒间隔重复40秒和25秒。我们在同一层面测量CT值,计算每个时间点的时间密度曲线(TDC)和TAG。

结果

在0%狭窄时,CO为2.0和4.0升/分钟时,沿TDC的每个时间点,TAG的变化比70%狭窄时小(2.0升/分钟时为-3.02至+0.55HU/cm,4.0升/分钟时为-2.63至+0.43HU/cm)。与70%狭窄且CO为2.0升/分钟相比,4.0升/分钟时CO的TAG曲线随时间逐渐变化(2.0升/分钟时为-6.64至+1.18HU/cm,4.0升/分钟时为-3.46至+2.75HU/cm)。

结论

尽管有无狭窄的连接管尺寸相同,但TAG值受CM注射后的扫描时间和CO影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验