Short TR, variable flip angle, gradient echo scans of the cervical spine: comparison of 2DFT and 3DFT techniques.

作者信息

Enzmann D, Rubin J B

机构信息

Department of Diagnostic Radiology and Nuclear Medicine, Stanford University School of Medicine, California.

出版信息

Neuroradiology. 1989;31(3):213-6. doi: 10.1007/BF00344345.

Abstract

A prospective study of 16 patients was performed to compare quantitatively a contiguous single slice 2DFT version with a 3DFT version of a short TR, variable flip angle, gradient echo (GRASS) pulse sequence. The 3DFT GRASS scans had higher signal-to-noise ratios (SNR) of cord and CSF compared to the single slice 2DFT GRASS scans. The 3DFT GRASS scans, however, had lower CSF-cord and CSF-disc contrast than the single slice 2DFT version. The 3DFT GRASS sequence demonstrated comparable contrast only on the end slices of an imaging volume suggesting influence of an entry phenomenon. The lower CSF-cord and CSF-disc contrast of the 3DFT GRASS technique diminished its usefulness in the diagnosis of cervical disc disease compared to the single slice 2DFT GRASS technique. Two different slice thicknesses (3 mm and 5 mm) were investigated with the 2DFT GRASS technique and found to be comparable although the 3 mm scans had sharper disc and dural margins because of less partial volume artifact.

摘要

对16例患者进行了一项前瞻性研究,以定量比较短TR、可变翻转角梯度回波(GRASS)脉冲序列的连续单层面2DFT版本与3DFT版本。与单层面2DFT GRASS扫描相比,3DFT GRASS扫描的脊髓和脑脊液具有更高的信噪比(SNR)。然而,3DFT GRASS扫描的脑脊液-脊髓和脑脊液-椎间盘对比度低于单层面2DFT版本。3DFT GRASS序列仅在成像容积的末端层面显示出相当的对比度,提示存在进入现象的影响。与单层面2DFT GRASS技术相比,3DFT GRASS技术较低的脑脊液-脊髓和脑脊液-椎间盘对比度降低了其在诊断颈椎间盘疾病中的效用。使用2DFT GRASS技术研究了两种不同的层厚(3mm和5mm),发现两者相当,尽管3mm扫描由于部分容积伪影较少而具有更清晰的椎间盘和硬脊膜边缘。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验