Yeatts R P, Crum B
Department of Ophthalmology, Bowman Gray School of Medicine, Wake Forest University Medical Center, Winston-Salem, NC.
Ophthalmic Surg. 1989 Jul;20(7):490-3.
A 4-mm segment of the tip of a 0000 probe broke off inside the nasolacrimal duct of an 11-month-old girl during probing for congenital nasolacrimal duct obstruction. Despite numerous attempts, the tip could not be extracted. Examination of the 000 end of the double-ended probe showed a narrow neck approximately 4 mm from the bulbous tip. Choosing an appropriate probe design, inspecting the probe before use, and selecting the appropriate probe size will help prevent this unusual complication.
在为一名11个月大女童探查先天性鼻泪管阻塞时,一根0000探针的尖端有4毫米长的一段在鼻泪管内折断。尽管多次尝试,折断的尖端仍无法取出。检查双端探针的000端发现,距球根状尖端约4毫米处有一个狭窄颈部。选择合适的探针设计、使用前检查探针以及选择合适的探针尺寸有助于预防这种罕见并发症。