一种测量人工眼活动度的新方法。

A Novel Method to Measure Artificial Eye Motility.

作者信息

Mourits Daphne L, Hartong Dyonne T, van Beek Johannes H M, Witte Birgit I, Tan H Stevie, Moll Annette C

机构信息

Departments of *Ophthalmology, †Ear & Hearing, and ‡Epidemiology & Biostatistics, VU University Medical Center, Amsterdam, The Netherlands.

出版信息

Ophthalmic Plast Reconstr Surg. 2017 Nov/Dec;33(6):413-418. doi: 10.1097/IOP.0000000000000804.

Abstract

PURPOSE

To measure objectively, reproducibly, and noninvasively artificial eye motility, a fundamental aspect in the (cosmetic) outcome of enucleation.

METHOD

A gaze- and pupil-tracking system, the "iView X," was implemented to measure horizontal eye motility. The system, with adjusted software, was tested with patients wearing 1 prosthetic eye after enucleation for retinoblastoma. Measurements were repeated 5 times in every patient and analyses were performed twice by 2 independent observers. Reproducibility was tested via linear mixed models. After the implementation of the method, more data were obtained, including more patients with a history of enucleation for the treatment of retinoblastoma for eye tracker measurement, and differences in motility percentages between implant types and sizes were analyzed via Mann-Whitney U tests.

RESULTS

The intraclass correlation coefficient of the interobserver variable in the patient test group (n = 27, 6-53 years) was 0.98 and 0.96 for measurement of left gaze and right gaze, respectively. Intraobserver variation was <0.001. In the total of 58 included patients for comparative analysis, mean difference of prosthesis motility compared with the contralateral eye in abduction was 57.1% (range 3.2-91.5%); in adduction 65.8% (range 24.0-92.0%). No statistical differences were found between the different implant types and sizes.

CONCLUSION

Motility measurements of the prosthetic eye in comparison to the contralateral eye using the iView X system are reproducible and reliable. This is, to the authors' knowledge, the first easy applicable, noninvasive, reproducible, and commercially available instrument to evaluate prosthesis motility. With the adjusted software program (freely available on request), a similar objective measurement can be performed worldwide. The implant size or type did not influence the outcome of the motility measurement; this finding deserves additional study.

摘要

目的

客观、可重复且无创地测量人工眼的运动,这是眼球摘除术(美容)效果的一个基本方面。

方法

采用一种注视和瞳孔追踪系统“iView X”来测量水平眼动。该系统通过调整软件,对视网膜母细胞瘤眼球摘除术后佩戴1只假眼的患者进行测试。每位患者重复测量5次,由2名独立观察者进行2次分析。通过线性混合模型测试可重复性。该方法实施后,获得了更多数据,包括更多有视网膜母细胞瘤眼球摘除治疗史的患者进行眼动追踪测量,通过Mann-Whitney U检验分析植入物类型和尺寸之间的运动百分比差异。

结果

患者测试组(n = 27,6 - 53岁)中观察者间变量的组内相关系数,左注视测量为0.98,右注视测量为0.96。观察者内变异<0.001。在纳入进行比较分析的58例患者中,假眼外展时与对侧眼相比的平均运动差异为57.1%(范围3.2 - 91.5%);内收时为65.8%(范围24.0 - 92.0%)。不同植入物类型和尺寸之间未发现统计学差异。

结论

使用iView X系统测量假眼与对侧眼相比的运动是可重复且可靠的。据作者所知,这是首个易于应用、无创、可重复且商业化的评估假眼运动的仪器。通过调整后的软件程序(可根据要求免费获取),全球范围内均可进行类似的客观测量。植入物的尺寸或类型不影响运动测量结果;这一发现值得进一步研究。