Gopal Bhavesh P, Bell Stephen, Soliman Noha M, Sia David I T, Stacey Andrew W, Carpenter David, Sagoo Mandeep S
Ocular Prosthetics Department, Moorfields Eye Hospital, London, UK.
Ocupeye Ltd, Warwick, UK.
Eye (Lond). 2025 Jul 26. doi: 10.1038/s41433-025-03936-6.
To compare standard rule measurement (gold standard, analogue) method of orbital implant motility against digital image analysis on smart phone measurement software, Medigrid and ImageJ.
Single centre observational, non-randomised, non-masked cohort study in enucleated patients.
A standard rule measurement (gold standard) was used to measure vertical and horizontal movements of the fellow eye, prosthesis, and orbital implant. At the same time, photographs taken in the same positions of gaze were analysed using ImageJ software (NIH, Bethesda, Maryland, USA) and the smartphone app mediGrid (IRISS Medical Technologies, UK). Bland-Altman plots were used to compare the digital modalities, with the current gold standard.
A total of 54 patients were tested for fellow eye motility, ocular prosthesis and orbital implant motility following primary enucleation. The median time since enucleation was 282 days (range: 45-11,278). Image-J had consistently lower limits of bias and agreement than mediGrid, suggesting that image-J is more equivalent to the standard rule measurement. The largest difference in the limit of agreement between mediGrid and ImageJ can be found most in upgaze and downgaze.
In our analyses, Image-J was found to be a more accurate measure of orbital implant motility, than mediGrid.
比较眼眶植入物活动度的标准直尺测量法(金标准,模拟法)与智能手机测量软件Medigrid和ImageJ上的数字图像分析法。
对眼球摘除患者进行单中心观察性、非随机、非盲队列研究。
采用标准直尺测量法(金标准)测量健眼、义眼和眼眶植入物的垂直和水平运动。同时,使用ImageJ软件(美国国立卫生研究院,马里兰州贝塞斯达)和智能手机应用程序mediGrid(英国IRISS医疗技术公司)对在相同注视位置拍摄的照片进行分析。使用Bland-Altman图将数字测量方法与当前的金标准进行比较。
共对54例初次眼球摘除术后的健眼活动度、眼假体和眼眶植入物活动度进行了测试。眼球摘除后的中位时间为282天(范围:45 - 11278天)。Image-J的偏差和一致性下限始终低于mediGrid,表明Image-J与标准直尺测量法更为等效。mediGrid和ImageJ之间一致性界限的最大差异多见于上视和下视时。
在我们的分析中,发现Image-J比mediGrid更能准确测量眼眶植入物的活动度。