Department of Instrumental Analytical Chemistry, University of Duisburg-Essen (UDE), Universitätsstraße 5, 45141 Essen, Germany.
Department of Sensors and Measurement Technology, Institute of Electrical Engineering and Measurement Technology, Leibniz Universität Hannover, Appelstr. 9A, 30167 Hannover, Germany.
Talanta. 2017 Jan 1;162:159-166. doi: 10.1016/j.talanta.2016.10.024. Epub 2016 Oct 4.
X-ray was utilized as an ionization source for differential ion mobility spectrometry (DMS) for the first time. The utilization of this ionization source increases the potential of DMS system for on-site based applications. The influence of experimental parameters (e.g. accelerating voltage, filament current, and separation field) on the analysis of model compounds was investigated and discussed. It was found that both the positive and the negative reactive ion peaks [RIP(+) and RIP(-)] formed during X-ray ionization are identical with those observed with the traditional Ni radioactive ion source. This is especially notable for RIP(-), because the chemistry provided by other nonradioactive sources in the negative mode is more complicated or even different than that observed with a Ni source. Increase of either filament current or accelerating voltage resulted in increased intensity of both RIP(+) and RIP(-). However, because of the materials used for construction of X-ray adapter the maximal level of filament current and accelerating voltage used in this study were limited to 700mA and 5kV, respectively. Analytical performance was determined with two model compounds (acetone and methyl salicylate) using X-ray and directly compared to Ni ionization source. When X-ray was coupled to DMS, calculated LOD values were found to be within the range of 0.17-1.52ppbv/v (concentration in the carrier gas). These values are competitive with those calculated for DMS equipped with traditional Ni radioactive ionization source. The obtained results are promising enough to ensure the potential of X-ray as ionization source for DMS.
X 射线首次被用作差分离子迁移谱(DMS)的电离源。这种电离源的使用增加了 DMS 系统用于现场应用的潜力。研究并讨论了实验参数(例如加速电压、灯丝电流和分离场)对模型化合物分析的影响。结果发现,X 射线电离过程中形成的正、负离子反应峰[RIP(+)和 RIP(-)]与传统的 Ni 放射性离子源观察到的相同。这对于 RIP(-)尤其明显,因为在负离子模式下,其他非放射性源提供的化学物质比使用 Ni 源观察到的更复杂,甚至不同。灯丝电流或加速电压的增加都会导致 RIP(+)和 RIP(-)的强度增加。然而,由于 X 射线适配器的结构材料,本研究中使用的灯丝电流和加速电压的最大水平分别限制在 700mA 和 5kV。使用 X 射线和 Ni 电离源对两种模型化合物(丙酮和甲基水杨酸酯)进行了分析性能测定,并直接进行了比较。当 X 射线与 DMS 耦合时,计算出的 LOD 值在 0.17-1.52ppbv/v(载气中的浓度)范围内。这些值与配备传统 Ni 放射性电离源的 DMS 计算出的值相当。所获得的结果非常有希望,足以确保 X 射线作为 DMS 电离源的潜力。