Abstract

PURPOSE

To assess short-term stability and conformational changes of the Nellix EndoVascular Aneurysm Sealing (EVAS) System using 3-dimensional (3D) analysis.

METHODS

Postoperative computed tomography (CT) scans obtained at 0, 3, and 12 months in 24 patients (mean age 75±7 years; 22 men) who underwent EVAS between December 2013 and December 2014 for intact abdominal aortic aneurysm (within the instructions for use) were evaluated for stent-graft deviation in multiple planes using dedicated 3D analysis software. In addition, 2D analysis using an anatomically fixed reference landmark was performed to assess craniocaudal migration. Clinical and follow-up data of the patients were recorded and matched with results of the imaging analysis.

RESULTS

Overall stability of the Nellix endografts was promising. Relevant conformational changes in the majority of cases were limited to the iliac graft segment and were clinically benign in all cases. Conversely, the only deviation of the proximal stent-graft segment was found in a patient with type Ia endoleak. Additional 2D analysis found relevant (≥5 mm) caudal migration of the Nellix stent-graft in 6 patients, including the one with the type Ia endoleak. In 3 patients, 3D analysis demonstrated the absence of relevant conformational changes of the endografts despite caudal migration.

CONCLUSION

Overall stability of the separate EVAS stent-grafts is promising in the short term. Relevant conformational changes (stent-graft deviation) in the majority of cases were benign and confined to the iliac segment.

摘要

目的

使用三维(3D)分析评估Nellix血管内动脉瘤封堵(EVAS)系统的短期稳定性和构象变化。

方法

对2013年12月至2014年12月期间因完整腹主动脉瘤(在使用说明书范围内)接受EVAS治疗的24例患者(平均年龄75±7岁;22例男性)在术后0、3和12个月获得的计算机断层扫描(CT)图像,使用专用3D分析软件在多个平面评估支架移植物的偏差。此外,使用解剖学固定的参考标志进行二维分析以评估头尾方向的移位。记录患者的临床和随访数据,并与影像学分析结果进行匹配。

结果

Nellix腔内移植物的整体稳定性良好。大多数病例中相关的构象变化仅限于髂部移植物段,且所有病例在临床上均为良性。相反,仅在1例Ia型内漏患者中发现近端支架移植物段有偏差。额外的二维分析发现6例患者(包括1例Ia型内漏患者)的Nellix支架移植物有相关(≥5 mm)的尾端移位。在3例患者中,三维分析显示尽管有尾端移位,但腔内移植物无相关构象变化。

结论

单独的EVAS支架移植物在短期内整体稳定性良好。大多数病例中相关的构象变化(支架移植物偏差)是良性的,且局限于髂部段。