Yamagishi T, Sasada T, Maeda J, Kohno T, Yoshinaga K, Ogawa H, Matsuda Y, Ozaki M, Kusukawa R
Kaku Igaku. 1989 Feb;26(2):155-64.
Accuracy of the measurements of LV volumes and ejection fraction determined from multigated blood pool tomography (MGBPT) was tested in 23 patients with various heart diseases who underwent cineventriculography (CV). Preliminary phantom studies showed that a 42% threshold value was found to provide the best relationship between measured and actual volumes (r = 0.99, standard error of the estimate (SEE) = 4.7 ml). The patients studies were performed at 16 frames/cardiac cycle at 18 angles over a 180 degrees rotation for one minute per angle. Long-axis horizontal views were reconstructed. The numbers of LV voxel with counts above the threshold value were summed and multiplied by the known volume of a voxel. Measurements of LV volume (r = 0.82, SEE = 29 ml), end-systolic LV volume (r = 0.83, SEE = 25 ml), end-diastolic LV volume (r = 0.68, SEE = 33 ml), and ejection fraction (r = 0.73, SEE = 10%) determined from MGBPT correlated well with those determined by CV. However, the ratios (Y) of the difference of the MGBPT-volume from the CV-volume to the CV-volume were decreased progressively and negatively as the CV-volumes (X) were smaller (Y = 38-3,996/X). Thus, we conclude that the determinations of volume by MGBPT would lead to more errors with the smaller volumes.
在23例患有各种心脏病且接受了电影心室造影(CV)的患者中,测试了通过多门控心血池断层扫描(MGBPT)测定左心室容积和射血分数的测量准确性。初步的模型研究表明,发现42%的阈值能在测量容积与实际容积之间提供最佳关系(r = 0.99,估计标准误差(SEE)= 4.7毫升)。对患者的研究在每个心动周期16帧、180度旋转范围内18个角度、每个角度持续1分钟的条件下进行。重建长轴水平视图。将计数高于阈值的左心室体素数量相加,并乘以已知的体素容积。由MGBPT测定的左心室容积(r = 0.82,SEE = 29毫升)、收缩末期左心室容积(r = 0.83,SEE = 25毫升)、舒张末期左心室容积(r = 0.68,SEE = 33毫升)和射血分数(r = 0.73,SEE = 10%)与CV测定的结果具有良好的相关性。然而,随着CV容积(X)变小,MGBPT容积与CV容积之差与CV容积的比值(Y)逐渐减小且呈负相关(Y = 38 - 3,996/X)。因此,我们得出结论,MGBPT测定容积时,容积越小误差越大。