Bouwman R W, van Engen R E, Broeders M J M, den Heeten G J, Dance D R, Young K C, Veldkamp W J H
Dutch Reference Centre for Screening (LRCB), Radboud University Medical Centre, The Netherlands.
Dutch Reference Centre for Screening (LRCB), Radboud University Medical Centre, The Netherlands.
Phys Med. 2016 Dec;32(12):1559-1569. doi: 10.1016/j.ejmp.2016.11.109. Epub 2016 Nov 23.
In mammography, images are processed prior to display. Current methodologies based on physical image quality measurements are however not designed for the evaluation of processed images. Model observers (MO) might be suitable for this evaluation. The aim of this study was to investigate whether the non-pre-whitening (NPW) MO can be used to predict human observer performance in mammography-like images by including different aspects of the human visual system (HVS).
The correlation between human and NPW MO performance has been investigated for the detection of disk shaped objects in simulated white noise (WN) and clustered lumpy backgrounds (CLB), representing quantum noise limited and mammography-like images respectively. The images were scored by the MO and five human observers in a 2-alternative forced choice experiment.
For WN images it was found that the log likelihood ratio (R), which expresses the goodness of fit, was highest (0.44) for the NPW MO without addition of HVS aspects. For CLB the R improved from 0.46 to 0.65 with addition of HVS aspects. The correlation was affected by object size and background.
This study shows that by including aspects of the HVS, the performance of the NPW MO can be improved to better predict human observer performance. This demonstrates that the NPW MO has potential for image quality assessment. However, due to the dependencies found in the correlation, the NPW MO can only be used for image quality assessment for a limited range of object sizes and background variability.
在乳腺钼靶摄影中,图像在显示之前会进行处理。然而,目前基于物理图像质量测量的方法并非为评估处理后的图像而设计。模型观察者(MO)可能适用于这种评估。本研究的目的是调查非白化(NPW)MO是否可用于通过纳入人类视觉系统(HVS)的不同方面来预测乳腺钼靶摄影样图像中的人类观察者表现。
分别针对在模拟白噪声(WN)和聚集性块状背景(CLB)中检测盘状物体的情况,研究了人类与NPW MO表现之间的相关性,WN和CLB分别代表量子噪声受限图像和乳腺钼靶摄影样图像。在二项迫选实验中,由MO和五名人类观察者对图像进行评分。
对于WN图像,发现表示拟合优度的对数似然比(R)在未添加HVS方面时,NPW MO最高(0.44)。对于CLB,添加HVS方面后,R从0.46提高到0.65。相关性受物体大小和背景的影响。
本研究表明,通过纳入HVS的各个方面,NPW MO的表现可以得到改善,从而更好地预测人类观察者的表现。这表明NPW MO在图像质量评估方面具有潜力。然而,由于在相关性中发现的依赖性,NPW MO仅可用于有限范围的物体大小和背景变异性的图像质量评估。