植入式心脏节律装置的模拟消隐期

The Analog Blanking Period of Implantable Cardiac Rhythm Devices.

作者信息

Barold S Serge, Kucher Andreas

机构信息

Department of Medicine, University of Rochester School of Medicine and Dentistry, Rochester, New York.

Biotronik, Berlin, Germany.

出版信息

Pacing Clin Electrophysiol. 2017 Feb;40(2):115-127. doi: 10.1111/pace.12982. Epub 2017 Jan 17.

Abstract

BACKGROUND

Analog blanking periods (BPs) that hold down the display of electrograms (EGMs) in cardiac rhythm devices have received much less attention than the well-known digital BPs which do not influence the EGM display. In Biotronik devices (Biotronik GmbH, Berlin, Germany), when a paced event initiates an analog BP in one chamber (right atrium, right ventricle [RV], or left ventricle [LV]), an identical cross-chamber analog BP starts in the other two chambers.

METHODS

All clinical observations were recorded from patients with Biotronik devices. The effect of the atrial cross-chamber analog BP initiated by a ventricular paced event on the atrial EGM was studied in the records of seven patients with frequent paroxysmal atrial flutter to determine the effect of critically timed RV paced event (RVp) or LV paced event (LVp) upon the atrial EGM. The effect of atrial pacing triggering cross-chamber analog BPs in the RV and LV channels on the RV and LV EGMs was also investigated in cases of conducted supraventricular beats and ventricular premature complexes. The effect of a triggered LVp initiating a cross-chamber analog BP in the RV channel on the EGM of a sensed RV sensed event was evaluated in 10 cases. Simulation studies were also performed to verify the clinical observations.

RESULTS

Patients with atrial flutter showed intermittent truncation or deformity and even disappearance of the atrial signals due to an atrial cross-chamber analog BP initiated by RVp and/or LVp. Three patients demonstrated deformity of the signal shape of ventricular premature beats falling within a ventricular cross-chamber analog BP initiated by right atrial paced event (RAp). We found only one case of a supraventricular QRS complex trapped in a ventricular cross-chamber analog BP initiated by RAp. All the recordings of LVp triggering upon RVs revealed a variety of RV signal deformities occasionally with preservation of the terminal part of the RV signal. Simulation studies confirmed the effect of the analog BP on the atrial and the ventricular EGMs.

CONCLUSION

The analog BP of Biotronik devices may cause truncation or deformity of atrial and ventricular signals and the occasional disappearance of an atrial signal during atrial flutter. These effects must not be interpreted as device malfunction. In the clinical evaluation of the EGM curves, the effects of the analog BPs have to be carefully considered in order to avoid possible misinterpretation.

摘要

背景

心脏节律装置中抑制心电图(EGM)显示的模拟消隐期(BP)受到的关注远少于不影响EGM显示的著名数字BP。在百多力装置(百多力有限公司,德国柏林)中,当一个起搏事件在一个腔室(右心房、右心室[RV]或左心室[LV])引发模拟BP时,相同的跨腔室模拟BP会在另外两个腔室启动。

方法

所有临床观察均记录自使用百多力装置的患者。在7例频发阵发性心房扑动患者的记录中,研究了心室起搏事件引发的心房跨腔室模拟BP对心房EGM的影响,以确定关键定时的右心室起搏事件(RVp)或左心室起搏事件(LVp)对心房EGM的影响。在房室传导的室上性搏动和室性早搏的病例中,还研究了心房起搏触发RV和LV通道中的跨腔室模拟BP对RV和LV EGM的影响。在10例病例中评估了触发的LVp在RV通道中引发跨腔室模拟BP对感知到的RV感知事件的EGM的影响。还进行了模拟研究以验证临床观察结果。

结果

心房扑动患者显示,由于RVp和/或LVp引发的心房跨腔室模拟BP,心房信号出现间歇性截断、变形甚至消失。3例患者表现出落入由右心房起搏事件(RAp)引发的心室跨腔室模拟BP内的室性早搏信号形状变形。我们仅发现1例室上性QRS波群被困在由RAp引发的心室跨腔室模拟BP内的情况。所有LVp触发RV的记录均显示出各种RV信号变形,偶尔会保留RV信号的末端部分。模拟研究证实了模拟BP对心房和心室EGM的影响。

结论

百多力装置的模拟BP可能导致心房和心室信号截断或变形,以及心房扑动期间心房信号偶尔消失。这些影响不应被解释为装置故障。在EGM曲线的临床评估中,必须仔细考虑模拟BP的影响,以避免可能的错误解读。