Bockisch A, Ruhlmann J, Hotze A, Biersack H J
Institut für klinische und experimentelle Nuklearmedizin, Universität Bonn, BRD.
Nuklearmedizin. 1989 Aug;28(4):145-7.
The attenuation effect of the female breast on single-photon emission computed 201Tl-chloride myocardial tomograms (SPECT) was quantitatively evaluated in a group of 20 women. SPECT was carried out with the left breast in the normal position and after its maximal upward displacement. The count rate gain was up to 30% when the breast was raised. 201Tl-chloride myocardial SPECT often showed false lesions in the anterior wall which resulted from the attenuation of the gamma rays by the female breast. This factor is neglected in constructing the tomograms, and even if such artifacts are absent, the attenuation may adversely affect the calculated wash-out values.
在一组20名女性中对女性乳房对单光子发射计算机断层扫描(SPECT)201Tl-氯化物心肌断层图像的衰减效应进行了定量评估。SPECT分别在左乳房处于正常位置时以及将其最大程度向上移位后进行。当乳房抬起时,计数率增益高达30%。201Tl-氯化物心肌SPECT在前壁经常显示出假病变,这是由女性乳房对γ射线的衰减所致。在构建断层图像时,这一因素被忽略了,并且即使不存在此类伪影,衰减也可能对计算出的洗脱值产生不利影响。