Roda A, Girotti S, Ghini S, Carrea G
Institute of Analytical Chemistry, University of Messina, Salita Sperone, Italy.
J Biolumin Chemilumin. 1989 Jul;4(1):423-35. doi: 10.1002/bio.1170040157.
Immobilized enzymes are widely used in the clinical laboratory in the assay of several analytes and enzymes. The use of immobilized enzymes makes these reagents recoverable and re-usable, and in most cases increases their stability and catalytic activity. In conjunction with bioluminescent enzymes (firefly and bacterial luciferases) and chemiluminescent catalyst (peroxidase) we set up high-sensitive flow methods based on the use of nylon tube coil or epoxy methacrylate column as solid support. All the NAD(P)/NAD(P)H-dependent dehydrogenases (bacterial luciferase), ATP-dependent enzymes (firefly luciferase) and oxidases producing H2O2 (peroxidase) can be immobilized and a large variety of analytes have been sensitively measured. As an alternative format we also reported a dry chemistry method in which all the enzymes, substrates and cofactors are ready to use, supported on dry cellulose disks. Methodological problems such as flow conditions, stability, pH, ionic strength and analytical performances are also reported.
固定化酶在临床实验室中广泛用于多种分析物和酶的检测。固定化酶的使用使这些试剂可回收和重复使用,并且在大多数情况下提高了它们的稳定性和催化活性。结合生物发光酶(萤火虫和细菌荧光素酶)和化学发光催化剂(过氧化物酶),我们基于使用尼龙管线圈或环氧甲基丙烯酸酯柱作为固体支持物建立了高灵敏度流动方法。所有依赖NAD(P)/NAD(P)H的脱氢酶(细菌荧光素酶)、依赖ATP的酶(萤火虫荧光素酶)以及产生H2O2的氧化酶(过氧化物酶)都可以被固定化,并且已经能够灵敏地检测多种分析物。作为另一种形式,我们还报道了一种干化学方法,其中所有的酶、底物和辅助因子都已准备好使用,负载在干燥的纤维素圆盘上。还报道了诸如流动条件、稳定性、pH、离子强度和分析性能等方法学问题。