Impact of thoracic endovascular aortic repair on radial strain in an ex vivo porcine model.

作者信息

Nauta Foeke J H, de Beaufort Hector W L, Conti Michele, Marconi Stefania, Kamman Arnoud V, Ferrara Anna, van Herwaarden Joost A, Moll Frans L, Auricchio Ferdinando, Trimarchi Santi

机构信息

Thoracic Aortic Research Center, Policlinico San Donato IRCCS, University of Milan, Italy.

Department of Vascular Surgery, University Medical Center Utrecht, The Netherlands.

出版信息

Eur J Cardiothorac Surg. 2017 Apr 1;51(4):783-789. doi: 10.1093/ejcts/ezw393.

Abstract

OBJECTIVES

To quantify the impact of thoracic endovascular aortic repair (TEVAR) on radial aortic strain with the aim of elucidating stent-graft-induced stiffening and complications.

METHODS

Twenty fresh thoracic porcine aortas were connected to a mock circulatory loop driven by a centrifugal flow pump. A high-definition camera captured diameters at five different pressure levels (100, 120, 140, 160, and 180 mmHg), before and after TEVAR. Three oversizing groups were created: 0-9% ( n  = 7), 10-19% ( n  = 6), and 20-29% ( n  = 6). Radial strain (or deformation) derived from diameter amplitude divided by baseline diameter at 100 mmHg. Uniaxial tensile testing evaluated Young's moduli of the specimens.

RESULTS

Radial strain was reduced after TEVAR within the stented segment by 49.4 ± 24.0% ( P  < 0.001). As result, a strain mismatch was observed between the stented segment and the proximal non-stented segment (7.0 ± 2.5% vs 11.8 ± 3.9%, P  < 0.001), whereas the distal non-stented segment was unaffected ( P  = 0.99). Stent-graft oversizing did not significantly affect the amount of strain reduction ( P  = 0.30). Tensile testing showed that the thoracic aortas tended to be more elastic proximally than distally ( P  = 0.11).

CONCLUSIONS

TEVAR stiffened the thoracic aorta by 2-fold. Such segmental stiffening may diminish the Windkessel function considerably and might be associated with TEVAR-related complications, including stent-graft-induced dissection and aneurysmal dilatation. These data may have implications for future stent-graft design, in particular for TEVAR of the highly compliant proximal thoracic aorta.

摘要

目的

量化胸主动脉腔内修复术(TEVAR)对主动脉径向应变的影响,以阐明支架移植物引起的硬化及并发症。

方法

将20个新鲜的猪胸主动脉连接到由离心式流量泵驱动的模拟循环回路。在TEVAR术前和术后,通过高清摄像头在五个不同压力水平(100、120、140、160和180 mmHg)下采集直径数据。创建了三个不同尺寸过大比例的组:0 - 9%(n = 7)、10 - 19%(n = 6)和20 - 29%(n = 6)。径向应变(或变形)通过直径振幅除以100 mmHg时的基线直径得出。单轴拉伸试验评估标本的杨氏模量。

结果

TEVAR术后,支架置入段的径向应变降低了49.4±24.0%(P < 0.001)。结果,在支架置入段和近端非支架置入段之间观察到应变不匹配(7.0±2.5%对11.8±3.9%,P < 0.001),而远端非支架置入段未受影响(P = 0.99)。支架移植物尺寸过大对应变降低量无显著影响(P = 0.30)。拉伸试验表明,胸主动脉近端比远端更具弹性(P = 0.11)。

结论

TEVAR使胸主动脉硬化了2倍。这种节段性硬化可能会显著削弱风箱功能,并可能与TEVAR相关并发症有关,包括支架移植物引起的夹层和动脉瘤扩张。这些数据可能对未来的支架移植物设计有启示,特别是对于高度顺应性的近端胸主动脉的TEVAR。