Stent-Graft Deployment Increases Aortic Stiffness in an Ex Vivo Porcine Model.

作者信息

de Beaufort Hector W L, Conti Michele, Kamman Arnoud V, Nauta Foeke J H, Lanzarone Ettore, Moll Frans L, van Herwaarden Joost A, Auricchio Ferdinando, Trimarchi Santi

机构信息

Thoracic Aortic Research Center, Policlinico San Donato IRCCS, University of Milan, Milan, Italy.

Department of Civil Engineering and Architecture, University of Pavia, Pavia, Italy.

出版信息

Ann Vasc Surg. 2017 Aug;43:302-308. doi: 10.1016/j.avsg.2017.04.024. Epub 2017 May 5.

Abstract

BACKGROUND

Aortic stiffness is an independent predictor of cardiovascular mortality. In this study, the effect of thoracic endovascular aortic repair (TEVAR) on aortic stiffness is investigated by measuring aortic pulse wave velocity (PWV) in an ex vivo porcine model.

METHODS

Fifteen fresh porcine thoracic aortas were connected to a benchtop pulsatile system. Intraluminal pressures were recorded in the ascending aorta and at the celiac trunk using a needle connected to a pressure sensor. The distance between the needles was divided by the time difference between the base of the pressure peaks to calculate aortic PWV at baseline and after stent-graft deployment and distal stent-graft extension.

RESULTS

Mean aortic PWV was 5.0 m/s at baseline. PWV increased by 4% after proximal stent-graft deployment (P = 0.09) and by 18% after stent-graft extension (P < 0.001). Pulse pressure in the nonstented ascending aorta increased by 11.0 ± 1.2 mm Hg after proximal stent-graft deployment (P < 0.001) and by 17.3 ± 1.5 mm Hg after stent-graft extension (P < 0.001). The increases in PWV and pulse pressure showed a positive linear correlation with the percentage of stent-graft coverage (P < 0.001 and P < 0.001).

CONCLUSIONS

In this experimental setup, aortic stiffness increased after stent-graft deployment, dependent on the percentage of the aorta that was covered by stent graft. These results show that TEVAR leads to significant changes in aortic hemodynamics, which merits evaluation in the clinical setting.

摘要

背景

主动脉僵硬度是心血管死亡率的独立预测因素。在本研究中,通过在体外猪模型中测量主动脉脉搏波速度(PWV),研究胸主动脉腔内修复术(TEVAR)对主动脉僵硬度的影响。

方法

将15个新鲜猪胸主动脉连接到台式搏动系统。使用连接到压力传感器的针头记录升主动脉和腹腔干处的腔内压力。用针头之间的距离除以压力峰值底部之间的时间差,以计算基线时、支架移植物展开后和远端支架移植物延伸后的主动脉PWV。

结果

基线时平均主动脉PWV为5.0m/s。近端支架移植物展开后PWV增加4%(P = 0.09),支架移植物延伸后增加18%(P < 0.001)。近端支架移植物展开后,未植入支架的升主动脉中的脉压增加11.0±1.2mmHg(P < 0.001),支架移植物延伸后增加17.3±1.5mmHg(P < 0.001)。PWV和脉压的增加与支架移植物覆盖百分比呈正线性相关(P < 0.001和P < 0.001)。

结论

在本实验装置中,支架移植物展开后主动脉僵硬度增加,这取决于主动脉被支架移植物覆盖的百分比。这些结果表明,TEVAR会导致主动脉血流动力学发生显著变化,这值得在临床环境中进行评估。