Hideji I, MacKeen D L, Hamano H, Jester J V, Cavanagh H D
Osaka University.
CLAO J. 1989 Oct-Dec;15(4):290-7.
We determined diurnal variation in corneal thickness in rabbits prior to and following overnight wear of: (i) selected rigid contact lenses with different Dk values; (ii) hydrogel lenses of low and high water content; and (iii) elastomer lenses. The degree of contact lens-induced corneal swelling observed during 24 hours of lens wear, and the rates of deswelling in the subsequent 24 hours, correlated well with the different oxygen transmissibilities of the individual RGP contact lenses. The greatest swelling (21.6 +/- 5.4%) followed the wear of PMMA lenses. The least swelling, 2.9 +/- 4.0%, followed the wear of rigid gas permeable (RGP) Menicon SF-P (melafocon A) lenses, a value nearly identical to the swelling observed in the morning following sleep without lenses, (0.0 +/- 3.1%). By contrast, low-water content hydrogel soft contact lens use was associated with drastic corneal deswelling rates (-15.1 +/- 4.5%) during the hours after lens wear. The difference between these and control corneas was significant by paired t-test (P less than 0.01). Eyes wearing high water content lenses had less deswelling than eyes with their low-water counterparts. Corneal swelling produced by elastomer lenses was similar to that seen with RGP lenses.
(i)具有不同Dk值的特定硬性隐形眼镜;(ii)低含水量和高含水量的水凝胶镜片;以及(iii)弹性体镜片。在佩戴镜片24小时期间观察到的隐形眼镜引起的角膜肿胀程度,以及随后24小时内的消肿速率,与各RGP隐形眼镜不同的透氧性密切相关。佩戴PMMA镜片后肿胀最为明显(21.6±5.4%)。肿胀程度最小的是佩戴硬性透气性(RGP)美尼康SF-P(美拉氟康A)镜片后,为2.9±4.0%,这一数值与未佩戴镜片睡眠后早晨观察到的肿胀程度(0.0±3.1%)几乎相同。相比之下,使用低含水量水凝胶软性隐形眼镜后,在佩戴镜片后的数小时内角膜消肿速率极快(-15.1±4.5%)。通过配对t检验,这些角膜与对照角膜之间的差异具有显著性(P<0.01)。佩戴高含水量镜片的眼睛消肿程度低于佩戴低含水量镜片的眼睛。弹性体镜片引起的角膜肿胀与RGP镜片相似。