Minami Keiichiro, Maruyama Yoko, Mihashi Toshifumi, Miyata Kazunori, Oshika Tetsuro
Miyata Eye Hospital, 6-3 Kurahara-cho, Miyakonojo, Miyazaki, 885-0051, Japan.
Department of Ophthalmology, Faculty of Medicine, University of Tsukuba, Tsukuba, Japan.
Jpn J Ophthalmol. 2017 Mar;61(2):189-194. doi: 10.1007/s10384-016-0495-4. Epub 2017 Jan 6.
To evaluate the influence of increases in light scattering on intraocular lens (IOL) surfaces on paraxial forward scattering using goniophotometry and Hartmann-Shack wavefront aberrometry.
Surface light scattering was reproduced experimentally by acceleratedly aging 4 intraocular lenses by 0, 3, 5, and 10 years each. Light scattering from both IOL surfaces was measured using Scheimpflug photography. The paraxial forward scattering from the aged IOLs was measured using a goniophotometer with a halogen light source (wavelength: 350-850 nm) and telecentric optics, and changes in the maximum intensity and full width at 10% of maximum intensity (FW10%) were evaluated. The influences on the retina image were examined using a Hartmann-Shack aberrometer (wavelength: 840 nm). The contrast and difference from the point spread function of the central centroids were evaluated.
The mean surface light scattering from both IOL surfaces ranged from 30.0 to 118.3 computer compatible tape (CCT) and increased with each aging year. Evaluations using the goniophotometer and the Hartmann-Shack aberrometer showed no significant change in the paraxial forward scattering with the aging year (P > .45, Kruskal-Wallis test), and no association with the surface light scattering intensity was found (P > .75, Spearman rank correlation).
This experimental study using aged IOLs demonstrated that surface light scattering does not influence paraxial forward scattering.
使用角度分光光度法和哈特曼-夏克波前像差仪评估人工晶状体(IOL)表面光散射增加对近轴前向散射的影响。
通过将4个人工晶状体分别加速老化0、3、5和10年,实验性地再现表面光散射。使用Scheimpflug摄影测量来自IOL两个表面的光散射。使用带有卤素光源(波长:350 - 850 nm)和远心光学系统的角度分光光度计测量老化IOL的近轴前向散射,并评估最大强度和最大强度的10%处的全宽(FW10%)的变化。使用哈特曼-夏克像差仪(波长:840 nm)检查对视网膜图像的影响。评估对比度以及与中心质心点扩散函数的差异。
IOL两个表面的平均表面光散射范围为30.0至118.3计算机兼容磁带(CCT),并且随老化年份增加。使用角度分光光度计和哈特曼-夏克像差仪的评估表明,近轴前向散射随老化年份无显著变化(P >.45,Kruskal-Wallis检验),并且未发现与表面光散射强度有关联(P >.75,Spearman等级相关性)。
这项使用老化IOL的实验研究表明,表面光散射不影响近轴前向散射。