在新生儿重症监护病房中使用小型移动数字成像系统降低辐射剂量并提高图像质量。

Reduced radiation dose and improved image quality using a mini mobile digital imaging system in a neonatal intensive care unit.

作者信息

Kim Tae-Hoon, Ryu Jong-Hyun, Jeong Chang-Won, Jun Hong Young, Heo Dong-Woon, Lee Seung Hyun, Oh Yeon Kyun, Lee Mi-Jung, Yoon Kwon-Ha

机构信息

Imaging Science Research Center, Wonkwang University School of Medicine, Iksan, Jeonbuk, South Korea.

Department of Pediatrics, Wonkwang University School of Medicine.

出版信息

Clin Imaging. 2017 Mar-Apr;42:165-171. doi: 10.1016/j.clinimag.2016.12.004. Epub 2016 Dec 6.

Abstract

This study was aimed to assess the radiation dose and image quality of a mini-mobile digital imaging (mini-DI) system for neonatal chest radiography and compared to conventional digital radiography (DR). A total of 64 neonates were examined and anatomical landmarks were assessed. The entrance surface dose of mini DI and conventional DR was 26.64±0.15 μGy and 49.11±1.46 μGy, respectively (p<0.001). The mean SNR values for mini-DI and DR were 233.2±5.1 and 31.6±1.2, and 10% MTF values were 131 and 161μm. A newly developed mini-DI is capable of preserving the diagnostic information with dose reduction in neonates under intensive care.

摘要

本研究旨在评估用于新生儿胸部X线摄影的微型移动数字成像(mini-DI)系统的辐射剂量和图像质量,并与传统数字X线摄影(DR)进行比较。共检查了64例新生儿,并评估了解剖标志。mini-DI和传统DR的体表入射剂量分别为26.64±0.15μGy和49.11±1.46μGy(p<0.001)。mini-DI和DR的平均信噪比分别为233.2±5.1和31.6±1.2,10%调制传递函数值分别为131μm和161μm。新开发的mini-DI能够在降低剂量的情况下保留重症监护新生儿的诊断信息。

相似文献

7
Dose-image quality study in digital chest radiography using Monte Carlo simulation.
Appl Radiat Isot. 2008 Sep;66(9):1213-7. doi: 10.1016/j.apradiso.2008.01.009. Epub 2008 Jan 26.