机械扩张式 Lotus 瓣膜与球囊扩张式 SAPIEN3 经导管心脏瓣膜的瓣膜性能比较:一项独立核心实验室分析的观察性研究。

Comparison of valve performance of the mechanically expanding Lotus and the balloon-expanded SAPIEN3 transcatheter heart valves: an observational study with independent core laboratory analysis.

机构信息

Department of cardiology, Thoraxcenter, Erasmus University Medical Center, Rotterdam, The Netherlands.

Cardialysis Clinical Trials Management and Core Laboratories, Rotterdam, The Netherlands.

出版信息

Eur Heart J Cardiovasc Imaging. 2018 Feb 1;19(2):157-167. doi: 10.1093/ehjci/jew280.

Abstract

AIMS

The Lotus and SAPIEN3 are second-generation transcatheter heart valves, which are designed to minimize paravalvular aortic regurgitation (PAR) after transcatheter aortic valve replacement (TAVR). We sought to compare both devices for valve performance and with emphasis on PAR by independent core laboratory analysis.

METHODS AND RESULTS

A total of 162 (79 Lotus and 83 SAPIEN3) consecutive patients (51% female, 80 ± 8 years, Logistic EuroSCORE 14.8 ± 9.4%) who underwent TAVR because of aortic stenosis were included. Patients with aortic valve-in-valve treatment were excluded. Pre-discharge echocardiograms were analysed by an independent core laboratory using the Valve Academic Research Consortium 2 criteria. There were no differences in baseline and procedural characteristics, except for a larger aortic annulus and sizing indices in SAPIEN3-treated patients and frequency of post-dilatation (0% in Lotus and 13.1% in SAPIEN3). Both valves have similar mean residual gradient, indexed effective orifice area and Doppler velocity index when adjusted to valve size. The frequency of mild (13.9% vs. 31.3%) and at least moderate (1.3% vs. 3.6%) PAR was less after Lotus than after SAPIEN3 implantation (P = 0.02). Multi-slice computed tomography-based annulus and left ventricular outflow tract diameters, calcification and percentage of oversizing were not different between those with or without mild or more PAR. On multivariate analysis, the use of Lotus valve was associated with less (odds ratio OR, 0.41, P = 0.03) occurrence of PAR.

CONCLUSION

Overall, haemodynamic performance was comparable between the Lotus and SAPIEN3 valves. Lotus valve required less oversizing and was associated with less PAR than SAPIEN3.

摘要

目的

Lotus 和 SAPIEN3 是第二代经导管心脏瓣膜,旨在最大限度地减少经导管主动脉瓣置换(TAVR)后瓣周主动脉瓣反流(PAR)。我们旨在通过独立核心实验室分析比较两种装置的瓣膜性能和 PAR 重点。

方法和结果

共纳入 162 例(79 例 Lotus 和 83 例 SAPIEN3)连续接受 TAVR 治疗的主动脉瓣狭窄患者(51%为女性,80±8 岁,Logistic EuroSCORE 14.8±9.4%)。排除主动脉瓣中瓣治疗的患者。使用独立核心实验室,根据 Valve Academic Research Consortium 2 标准分析出院前超声心动图。两组患者在基线和手术特征方面无差异,除 SAPIEN3 治疗患者的主动脉瓣环和大小指数较大以及后扩张频率(Lotus 为 0%,SAPIEN3 为 13.1%)较高外。当调整到瓣膜大小后,两种瓣膜的平均残余梯度、指数有效瓣口面积和多普勒速度指数相似。Lotus 组轻度(13.9%比 31.3%)和至少中度(1.3%比 3.6%)PAR 的发生率低于 SAPIEN3 组(P=0.02)。轻度或更严重 PAR 患者与无 PAR 患者的多层螺旋 CT 基于瓣环和左心室流出道直径、钙化和过度放大百分比无差异。多变量分析显示,使用 Lotus 瓣膜与 PAR 发生率较低相关(比值比 OR,0.41,P=0.03)。

结论

总体而言,Lotus 和 SAPIEN3 瓣膜的血流动力学性能相当。Lotus 瓣膜需要的过度放大较少,与 SAPIEN3 相比 PAR 发生率较低。