Abstract

PURPOSE

The purpose of this work was to assess the impact of respiratory motion and to compare methods for suppression of respiratory motion artifacts in 4D Flow MRI.

METHODS

A numerical 3D aorta phantom was designed based on an aorta velocity field obtained by computational fluid mechanics. Motion-distorted 4D Flow MRI measurements were simulated and several different motion-suppression techniques were evaluated: Gating with fixed acceptance window size, gating with different window sizes in inner and outer k-space, and k-space reordering. Additionally, different spatial resolutions were simulated.

RESULTS

Respiratory motion reduced the image quality. All motion-suppression techniques improved the data quality. Flow rate errors of up to 30% without gating could be reduced to less than 2.5% with the most successful motion suppression methods. Weighted gating and gating combined with k-space reordering were advantageous compared with conventional fixed-window gating. Spatial resolutions finer than the amount of accepted motion did not lead to improved results.

CONCLUSION

Respiratory motion affects 4D Flow MRI data. Several different motion suppression techniques exist that are capable of reducing the errors associated with respiratory motion. Spatial resolutions finer than the degree of accepted respiratory motion do not result in improved data quality. Magn Reson Med 78:1877-1882, 2017. © 2017 International Society for Magnetic Resonance in Medicine.

摘要

目的

本研究旨在评估呼吸运动的影响,并比较在 4D Flow MRI 中抑制呼吸运动伪影的方法。

方法

基于计算流体力学获得的主动脉速度场,设计了一个数值 3D 主动脉体模。模拟了运动失真的 4D Flow MRI 测量,并评估了几种不同的运动抑制技术:固定接受窗大小的门控、内外 K 空间不同窗大小的门控和 K 空间重排。此外,还模拟了不同的空间分辨率。

结果

呼吸运动降低了图像质量。所有运动抑制技术都改善了数据质量。在没有门控的情况下,流速误差高达 30%,而最成功的运动抑制方法可将其降低至 2.5%以下。加权门控和门控结合 K 空间重排与传统的固定窗门控相比具有优势。空间分辨率比接受运动的程度更精细并不会导致结果的改善。

结论

呼吸运动影响 4D Flow MRI 数据。存在几种不同的运动抑制技术,能够降低与呼吸运动相关的误差。空间分辨率比接受呼吸运动的程度更精细不会导致数据质量的改善。磁共振医学 78:1877-1882, 2017。© 2017 国际磁共振学会。