Vyskocil Erich, Liepins Rudolfs, Kaider Alexandra, Blineder Michaela, Hamzavi Sasan
*Department of Otorhinolaryngology †Center for Medical Statistics, Informatics, and Intelligent Systems, Medical University of Vienna, Vienna, Austria.
Otol Neurotol. 2017 Mar;38(3):318-324. doi: 10.1097/MAO.0000000000001328.
There is no consensus regarding the benefit of implantable hearing aids in congenital unilateral conductive hearing loss (UCHL). This study aimed to measure sound source localization performance in patients with congenital UCHL and contralateral normal hearing who received a new bone conduction implant.
Evaluation of within-subject performance differences for sound source localization in a horizontal plane.
Tertiary referral center.
Five patients with atresia of the external auditory canal and contralateral normal hearing implanted with transcutaneous bone conduction implant at the Medical University of Vienna were tested.
Activated/deactivated implant.
Sound source localization test; localization performance quantified using the root mean square (RMS) error.
Sound source localization ability was highly variable among individual subjects, with RMS errors ranging from 21 to 40 degrees. Horizontal plane localization performance in aided conditions showed statistically significant improvement compared with the unaided conditions, with RMS errors ranging from 17 to 27 degrees. The mean RMS error decreased by a factor of 0.71 (p < 0.001).
Analysis revealed improved sound localization performance in a horizontal plane with the activated transcutaneous bone conduction implant. Some patients with congenital UCHL might be capable of developing improved horizontal plane localization abilities with the binaural cues provided by this device.
关于植入式助听器对先天性单侧传导性听力损失(UCHL)的益处尚无共识。本研究旨在测量接受新型骨传导植入物的先天性UCHL且对侧听力正常的患者的声源定位性能。
评估受试者在水平面内声源定位的性能差异。
三级转诊中心。
对维也纳医科大学5例外耳道闭锁且对侧听力正常并植入经皮骨传导植入物的患者进行测试。
激活/停用植入物。
声源定位测试;使用均方根(RMS)误差量化定位性能。
个体受试者之间的声源定位能力差异很大,RMS误差范围为21至40度。与未使用辅助装置的情况相比,使用辅助装置时在水平面内的定位性能有统计学上的显著改善,RMS误差范围为17至27度。平均RMS误差降低了0.71倍(p<0.001)。
分析显示激活的经皮骨传导植入物在水平面内的声音定位性能有所改善。一些先天性UCHL患者可能能够借助该装置提供的双耳线索提高水平面内的定位能力。