This study aims to evaluate the 2-year post-operative clinical outcomes of patients undergoing total knee arthroplasty (TKA) with computer-assisted surgery (CAS) using the pinless BrainLAB VectorVision Knee 2.5 Navigation System versus standard CAS. A retrospective case-control study of 200 patients who underwent TKA with CAS from 2008 to 2012 was conducted. Patients in the pinless arm underwent a shorter duration of surgery compared to the standard CAS arm with 72 ± 13 min and 83 ± 11 min, respectively ( p < 0.001), with improvement in Oxford Knee Score from 34 ± 8 to 18 ± 5 and 34 ± 9 to 18 ± 5, respectively, without statistically significant differences in clinical outcomes at 2 years.

摘要

本研究旨在评估使用无针BrainLAB VectorVision Knee 2.5导航系统的计算机辅助手术(CAS)与标准CAS进行全膝关节置换术(TKA)患者的术后2年临床结局。对2008年至2012年接受CAS-TKA的200例患者进行了一项回顾性病例对照研究。与标准CAS组相比,无针组手术时间更短,分别为72±13分钟和83±11分钟(p<0.001),牛津膝关节评分分别从34±8改善至18±5和从34±9改善至18±5,2年时临床结局无统计学显著差异。