调频 bSSFP 的多频重建。

Multifrequency reconstruction for frequency-modulated bSSFP.

机构信息

Department of Diagnostic and Interventional Radiology, University of Würzburg, Würzburg, Germany.

Department of Nuclear Medicine, University of Würzburg, Würzburg, Germany.

出版信息

Magn Reson Med. 2017 Dec;78(6):2226-2235. doi: 10.1002/mrm.26630. Epub 2017 Feb 10.

Abstract

PURPOSE

Banding artifacts in images acquired by balanced steady-state free precession (bSSFP) remain a challenge in MRI as they considerably reduce image quality, and diagnostic value deteriorates accordingly. As the steady-state tolerates small shifts in frequency, it is possible to acquire frequency-modulated bSSFP. Unfortunately, standard reconstructions of such measurements suffer from signal loss. Our study proposes a multifrequency reconstruction and demonstrates its capability of suppressing banding artifacts while retaining the high signal level of standard bSSFP.

METHODS

Numerical simulations in vitro and in vivo measurements were performed using both standard bSSFP and frequency-modulated bSSFP. The modulated data were reconstructed using a multifrequency approach consisting of three steps: phase correction, multiple reconstructions for different assumed frequencies, and maximum intensity projection.

RESULTS

Although standard bSSFP measurements showed banding artifacts that compromised the image quality, standard reconstructions of frequency-modulated acquisitions suffered from signal loss. In contrast, images reconstructed from frequency-modulated data using the proposed multifrequency reconstruction showed no visual bandings and featured a higher signal-to-noise ratio (SNR). The SNR gain for phantom and in vivo measurements ranged from 1.23 to 1.49.

CONCLUSIONS

The presented multifrequency reconstruction for frequency-modulated bSSFP provides images showing no bandings and featuring high SNR in short scan times. Magn Reson Med 78:2226-2235, 2017. © 2017 International Society for Magnetic Resonance in Medicine.

摘要

目的

在 MRI 中,平衡稳态自由进动(bSSFP)采集的图像中的带纹伪影仍然是一个挑战,因为它们会极大地降低图像质量,从而降低诊断价值。由于稳态可以容忍频率的微小偏移,因此可以采集频率调制的 bSSFP。不幸的是,这种测量的标准重建会有信号损失。本研究提出了一种多频率重建方法,并证明了其在抑制带纹伪影的同时保留标准 bSSFP 高信号水平的能力。

方法

在体外和体内分别使用标准 bSSFP 和频率调制 bSSFP 进行数值模拟和测量。使用包括三个步骤的多频率方法对调制数据进行重建:相位校正、针对不同假设频率的多次重建以及最大强度投影。

结果

尽管标准 bSSFP 测量显示带纹伪影会影响图像质量,但频率调制采集的标准重建会有信号损失。相比之下,使用提出的多频率重建从频率调制数据重建的图像没有可见的带纹,并且具有更高的信噪比(SNR)。对于体模和体内测量,SNR 增益范围为 1.23 至 1.49。

结论

本文提出的用于频率调制 bSSFP 的多频率重建方法可以在短扫描时间内提供无带纹且 SNR 高的图像。磁共振医学 78:2226-2235,2017。© 2017 国际磁共振学会。