影响口腔内扫描边缘质量的混杂因素。

Confounding factors affecting the marginal quality of an intra-oral scan.

机构信息

School of Dentistry, University of Leeds, Leeds LS2 9LU, United Kingdom.

School of Dentistry, University of Leeds, Leeds LS2 9LU, United Kingdom.

出版信息

J Dent. 2017 Apr;59:33-40. doi: 10.1016/j.jdent.2017.02.003. Epub 2017 Feb 9.

Abstract

OBJECTIVES

To assess the effect of clinical factors on the quality of intra-oral scans of crown margins. These factors are; presence of adjacent teeth, proximity to gingivae, encumbrance of wand positioning within oral cavity.

METHODS

A typodont lower molar (Frasaco, Germany) was prepared for an all-ceramic crown with 1.5mm supragingival (lingual) and equigingival (buccal) margins. The tooth was scanned in a model scanner, creating a master scan. An intra-oral scanner (IOS) (Omnicam, Sirona Dental) was used to acquire sets of 5 scans each, under varying conditions; (1) the presence/absence of adjacent teeth, (2) model mounted in manikin head/hand-held, (3) with/without a 1mm shim to elevate the margin. Every combination was investigated, yielding 40 scans (8 groups of 5). The master scan margin was identified by selecting the highest curvature region (>1.8). The master was aligned to each IOS scan, and 4 regions of each IOS scan margin were extracted, lying within 100μm of predefined mesial, distal, buccal and lingual sections of the master margin. The mean curvature of each margin section was calculated using Meshlab. The effect of each confounding factor on margin curvature was analysed using ANOVA.

RESULTS

Lingual margin curvature remained consistent regardless of scanning conditions. Buccal margin curvature was significantly affected when located equigingivally. Mesial margin curvature was significantly affected in the presence of adjacent teeth and proximity to the gingivae. Distal margin curvature was significantly affected by all three confounding factors.

CONCLUSIONS

The curvature (sharpness) of the margin recorded by a commercial IOS is significantly affected by clinical factors obscuring visibility.

摘要

目的

评估临床因素对牙冠边缘口内扫描质量的影响。这些因素包括:邻牙的存在、与牙龈的接近程度、口腔内扫描器探头定位的阻碍。

方法

在 Frasaco 德国的牙模下磨牙上制备全瓷冠,龈上(舌侧)和齐龈(颊侧)边缘为 1.5mm。牙在模型扫描仪中扫描,创建主扫描。使用口内扫描仪(IOS)(Sirona Dental 的 Omnicam)在不同条件下采集 5 个扫描集:(1)邻牙的存在/缺失,(2)模型安装在头模/手持,(3)使用/不使用 1mm 垫片抬高边缘。研究了每种组合,产生了 40 个扫描(8 组 5 个)。通过选择曲率最高的区域(>1.8)来识别主扫描边缘。将主扫描与每个 IOS 扫描对齐,并从每个 IOS 扫描边缘提取 4 个区域,位于主扫描边缘的预设近中、远中、颊侧和舌侧 100μm 内。使用 Meshlab 计算每个边缘区域的平均曲率。使用方差分析(ANOVA)分析每个混杂因素对边缘曲率的影响。

结果

无论扫描条件如何,舌侧边缘曲率保持一致。当位于齐龈时,颊侧边缘曲率受到显著影响。在存在邻牙和接近牙龈时,近中边缘曲率受到显著影响。所有三个混杂因素都显著影响远中边缘曲率。

结论

商业 IOS 记录的边缘曲率(尖锐度)显著受到遮蔽可见度的临床因素影响。