一种用于确定全膝关节置换术胫骨组件无菌性松动的新型基于百分比的系统。

A Novel Percentage-Based System for Determining Aseptic Loosening of Total Knee Arthroplasty Tibial Components.

作者信息

Chalmers Brian P, Sculco Peter K, Fehring Keith A, Trousdale Robert T, Taunton Michael J

机构信息

Department of Orthopedic Surgery, Mayo Clinic, Rochester, Minnesota.

出版信息

J Arthroplasty. 2017 Jul;32(7):2274-2278. doi: 10.1016/j.arth.2017.02.020. Epub 2017 Feb 16.

Abstract

BACKGROUND

There are limited data on evaluating the significance of radiolucent lines and aseptic loosening in total knee arthroplasty (TKA). We sought to compare the sensitivity, specificity, and reliability of the Knee Society Total Knee Arthroplasty Roentgenographic Evaluation and Scoring System (KSRES) in detecting tibial component loosening compared to a novel percentage-based system (PBS).

METHODS

We retrospectively reviewed radiographs obtained from 48 patients within 6 months prior to revision TKA. The radiographs were randomized and four reviewers independently used the KSRES to categorize tibial implants as nonconcerning, clinical follow-up for progression, or loose as described by KSRES. For the PBS, the percent involvement of the tibial implant interface of any radiolucency at the bone-cement or cement-implant interface was determined. The higher percentage from either the anteroposterior or the lateral image was the final score. Components were categorized as nonconcerning (≤10%), clinical follow-up for progression (11%-24%), or loose (≥25%). We compared the sensitivity, specificity, and interobserver reliability using intraoperative assessment of implant fixation as the gold standard.

RESULTS

For the KSRES, the mean sensitivity for determining tibial loosening was 7.3% and mean specificity for determining a nonconcerning implant was 95.9%. The PBS significantly increased the sensitivity to 91.1% (P < .001) while maintaining a specificity of 87.9% (P = .2). Interobserver reliability significantly increased from a mean kappa of 0.26 to 0.75 (P < .001).

CONCLUSION

The KSRES significantly underestimates implant loosening. The proposed percentage-based system demonstrated excellent sensitivity, specificity, and interobserver reliability in determining tibial implant loosening in this patient population.

摘要

背景

关于评估全膝关节置换术(TKA)中透亮线和无菌性松动的意义的数据有限。我们试图比较膝关节协会全膝关节置换术X线评估和评分系统(KSRES)与一种新的基于百分比的系统(PBS)在检测胫骨组件松动方面的敏感性、特异性和可靠性。

方法

我们回顾性分析了48例翻修TKA术前6个月内获得的X线片。将X线片随机分组,四位阅片者独立使用KSRES将胫骨植入物分类为无异常、临床随访观察进展情况或如KSRES所述为松动。对于PBS,确定在骨水泥或水泥 - 植入物界面处任何透亮区累及胫骨植入物界面的百分比。前后位或侧位图像中较高的百分比即为最终得分。组件被分类为无异常(≤10%)、临床随访观察进展情况(11% - 24%)或松动(≥25%)。我们以术中对植入物固定的评估作为金标准,比较了敏感性、特异性和观察者间可靠性。

结果

对于KSRES,确定胫骨松动的平均敏感性为7.3%,确定无异常植入物的平均特异性为95.9%。PBS显著提高了敏感性至91.1%(P < .001),同时保持特异性为87.9%(P = .2)。观察者间可靠性从平均kappa值0.26显著提高至0.75(P < .001)。

结论

KSRES显著低估了植入物松动情况。所提出的基于百分比的系统在确定该患者群体胫骨植入物松动方面表现出优异的敏感性、特异性和观察者间可靠性。