Yalin Serkan Feyyaz, Gulcicek Sibel, Avci Suna, Erkalma Senates Banu, Altiparmak Mehmet Riza, Trabulus Sinan, Alagoz Selma, Yavuzer Hakan, Doventas Alper, Seyahi Nurhan
Division of Nephrology, Department of Internal Medicine, Cerrahpasa Medical Faculty, Istanbul University, Istanbul, Turkey.
Division of Nephrology, Department of Internal Medicine, Istanbul Research and Training Hospital, Istanbul, Turkey.
Nephrology (Carlton). 2018 May;23(5):438-445. doi: 10.1111/nep.13042.
Bioelectrical impedance analysis is a promising method in determining the body compartments in haemodialysis patients. In this study, we aimed to investigate the agreement between two widely used methods: the single-frequency and multi-frequency bioelectrical impedance analyses.
Maintenance haemodialysis patients were enrolled in the study. Single-frequency and multi-frequency bioelectrical impedance analyses were performed consecutively before haemodialysis. A second bioelectrical impedance analysis was performed right after the haemodialysis session. A third bioelectrical impedance analysis was performed one hour after haemodialysis. We used weight change as a measure of fluid removal during haemodialysis session.
Bioelectrical impedance analysis estimates from both devices had significant differences. Best agreement was observed between single frequency and multifrequency devices' extracellular water estimates immediately after haemodialysis (mean difference 0.076 L). We found the best agreement between weight change and extracellular water change using single-frequency bioimpedance analysis. Moreover, one hour waiting time did not improve the agreement between weight and extracellular water changes for both devices. Different estimates seem to be caused by different raw impedance data measured by both devices and device-specific equations.
There are significant differences among bioelectrical impedance measurements performed with different bioelectrical impedance analyzers. Using open source software might be an important step forward in the development of standardized measurements.
生物电阻抗分析是一种用于确定血液透析患者身体组成部分的有前景的方法。在本研究中,我们旨在探究两种广泛使用的方法之间的一致性:单频生物电阻抗分析和多频生物电阻抗分析。
维持性血液透析患者被纳入本研究。在血液透析前连续进行单频和多频生物电阻抗分析。在血液透析结束后立即进行第二次生物电阻抗分析。在血液透析后一小时进行第三次生物电阻抗分析。我们使用体重变化作为血液透析过程中液体清除的指标。
两种设备的生物电阻抗分析估计值存在显著差异。血液透析后立即进行的单频和多频设备细胞外液估计值之间观察到最佳一致性(平均差异0.076升)。我们发现使用单频生物阻抗分析时体重变化与细胞外液变化之间的一致性最佳。此外,一小时的等待时间并未改善两种设备体重与细胞外液变化之间的一致性。不同的估计值似乎是由两种设备测量的不同原始阻抗数据和特定于设备的方程所导致的。
使用不同生物电阻抗分析仪进行的生物电阻抗测量之间存在显著差异。使用开源软件可能是标准化测量发展中的重要一步。