Sistani Habiballah, Karki Santosh, Archer Jieutonne J, Shi Fengjian, Levis Robert J
Department of Chemistry and Center for Advanced Photonics Research, Temple University, Philadelphia, PA, 19122, USA.
J Am Soc Mass Spectrom. 2017 May;28(5):880-886. doi: 10.1007/s13361-017-1622-2. Epub 2017 Mar 15.
A nonresonant, femtosecond (fs) laser is employed to desorb samples of Victoria blue deposited on stainless steel or indium tin oxide (ITO) slides using either electrospray deposition (ESD) or dried droplet deposition. The use of ESD resulted in uniform films of Victoria blue whereas the dried droplet method resulted in the formation of a ring pattern of the dye. Laser electrospray mass spectrometry (LEMS) measurements of the ESD-prepared films on either substrate were similar and revealed lower average relative standard deviations for measurements within-film (20.9%) and between-films (8.7%) in comparison to dried droplet (75.5% and 40.2%, respectively). The mass spectral response for ESD samples on both substrates was linear (R > 0.99), enabling quantitative measurements over the selected range of 7.0 × 10 to 2.8 × 10 mol, as opposed to the dried droplet samples where quantitation was not possible (R = 0.56). The limit of detection was measured to be 210 fmol. Graphical Abstract ᅟ.
使用非共振飞秒(fs)激光,通过电喷雾沉积(ESD)或干滴沉积法,解吸沉积在不锈钢或氧化铟锡(ITO)载玻片上的维多利亚蓝样品。采用ESD可得到均匀的维多利亚蓝薄膜,而干滴法则会形成染料的环状图案。对两种基底上ESD制备的薄膜进行激光电喷雾质谱(LEMS)测量结果相似,与干滴法相比(分别为75.5%和40.2%),其膜内测量(20.9%)和膜间测量(8.7%)的平均相对标准偏差更低。两种基底上ESD样品的质谱响应呈线性(R>0.99),能够在7.0×10至2.8×10摩尔的选定范围内进行定量测量,而干滴样品则无法进行定量(R = 0.56)。检测限测得为210飞摩尔。图形摘要ᅟ。