Suppr超能文献

用于实时质谱直接分析的替代质量参考标准品。

Alternative mass reference standards for direct analysis in real time mass spectrometry.

作者信息

Cody Robert B, Dane A John

机构信息

JEOL USA, Inc., 11 Dearborn Rd., Peabody, MA, 01960, USA.

出版信息

Rapid Commun Mass Spectrom. 2016 May 30;30(10):1206-1212. doi: 10.1002/rcm.7554.

Abstract

RATIONALE

Mass spectra were acquired with the Direct Analysis in Real Time (DART®) ion source for an amine-terminated polyether used as positive-ion mass reference standards and for several fluorinated materials commonly used as negative-ion reference standards for mass spectrometry.

METHODS

A commercial time-of-flight mass spectrometer equipped with a DART ion source was used for all measurements. Mass reference standards deposited onto the sealed end of a glass melting point tube were suspended in the DART gas stream for analysis.

RESULTS

A polyetheramine (Jeffamine® M-600) produced intense peaks corresponding to protonated molecules. Perfluorotributylamine (PFTBA), and perfluorotripentylamine, gave useful reference spectra for different m/z ranges. DART mass spectra of Ultramark 1621® resembled those previously reported for Fast Atom Bombardment (FAB) and Electrospray Ionization (ESI). Fomblin®Y, a fluorinated ether, was the most useful negative-ion reference standard of the materials tested. The material is commercially available, inexpensive, and provides reference peaks covering the m/z range 85 to >3000.

CONCLUSIONS

Jeffamine-M600 was found to be a convenient alternative to polyethers such as polyethylene glycol (PEG) for DART positive-ion mass calibration. Fomblin Y was suitable for use as a negative-ion reference standard. Copyright © 2016 John Wiley & Sons, Ltd.

摘要

原理

使用实时直接分析(DART®)离子源采集胺基封端聚醚(用作正离子质谱参考标准品)以及几种常用于质谱负离子参考标准品的含氟材料的质谱。

方法

所有测量均使用配备DART离子源的商用飞行时间质谱仪。将沉积在玻璃熔点管密封端的质谱参考标准品悬浮于DART气流中进行分析。

结果

一种聚醚胺(Jeffamine® M - 600)产生了与质子化分子相对应的强峰。全氟三丁胺(PFTBA)和全氟三戊胺在不同的m/z范围内给出了有用的参考光谱。Ultramark 1621®的DART质谱与先前报道的快原子轰击(FAB)和电喷雾电离(ESI)质谱相似。含氟醚Fomblin®Y是所测试材料中最有用的负离子参考标准品。该材料可商购,价格便宜,并提供覆盖m/z范围85至>3000的参考峰。

结论

发现Jeffamine - M600是用于DART正离子质谱校准的聚醚(如聚乙二醇(PEG))的便捷替代品。Fomblin Y适用于作为负离子参考标准品。版权所有© 2016 John Wiley & Sons, Ltd.