Gross Jürgen H
Eur J Mass Spectrom (Chichester). 2014;20(2):155-61. doi: 10.1255/ejms.1266.
A method for accurate mass calibration covering a wide m/z range in negative-ion direct analysis in real time mass spectrometry (DART-MS) is described. It is based on a commercially available tuning solution (Agilent ES tune mix) normally applied in electrospray ionization. This acetonitrile solution is composed of ammonium trifluoroacetate, betaine, 2,4,6-tris(heptafluoropropyl)-1,3,5-triazine, and various symmetrical hexakis-(fluoroalkoxy)-phosphazenes. Upon addition of 2% (v/vl trifluoroacetic acid in negative-ion DART-MS, this mixture yields negative ions covering a range of m/z 69 to m/z 2834. Validity of the calibration procedure is demonstrated on a Fourier transform ion cyclotron resonance mass spectrometer by application to polyethylene glycol diacid and [60]fullerene. Experimental conditions and a mass reference list are provided.
描述了一种在负离子实时直接分析质谱法(DART-MS)中实现宽m/z范围精确质量校准的方法。它基于通常用于电喷雾电离的市售调谐溶液(安捷伦ES调谐混合物)。这种乙腈溶液由三氟乙酸铵、甜菜碱、2,4,6-三(七氟丙基)-1,3,5-三嗪和各种对称的六(氟烷氧基)-磷腈组成。在负离子DART-MS中加入2%(v/v)三氟乙酸后,该混合物产生覆盖m/z 69至m/z 2834范围的负离子。通过应用于聚乙二醇二酸和[60]富勒烯,在傅里叶变换离子回旋共振质谱仪上证明了校准程序的有效性。提供了实验条件和质量参考列表。