新生儿颞动脉温度测量

Temporal Artery Temperature Measurement in the Neonate.

作者信息

Syrkin-Nikolau Mashette E, Johnson Karen J, Colaizy Tarah T, Schrock Ruthann, Bell Edward F

机构信息

Department of Pediatrics, University of Iowa, Iowa City, Iowa.

出版信息

Am J Perinatol. 2017 Aug;34(10):1026-1031. doi: 10.1055/s-0037-1601440. Epub 2017 Apr 10.

Abstract

We compared an infrared temporal artery thermometer with our clinical standard axillary thermometer for temperature measurements in neonatal patients.  We measured temporal artery (T), axillary (T, clinical standard), and rectal (T, gold standard) temperatures of 49 infants. The difference between T and T was compared with that between T and T, and the data were analyzed based on bed type and postmenstrual age.  The mean T, T, and T were 37.16 (SD 0.36) °C, 36.61 (SD 0.30) °C, and 36.82 (SD 0.30) °C, respectively. The measurements by these methods were all significantly different. The mean T-T was 0.21 (SD 0.26) °C, and the mean T-T was -0.34 (SD 0.37) °C, indicating that T was closer to T than was T ( < 0.0001). T agreed more closely with T for infants in cribs than for those in incubators. Adjusting for bed type and body weight, with each week of postmenstrual age, the discrepancy between T-T and T-T decreased by 0.005°C ( = 0.034).  Compared with the gold standard, T, T is not more accurate than T. The temporal artery thermometer was less accurate for infants in incubators than for infants in cribs. The accuracy of temporal artery temperature increased with postmenstrual age.

摘要

我们将一款红外颞动脉体温计与我们用于新生儿患者体温测量的临床标准腋温计进行了比较。我们测量了49名婴儿的颞动脉温度(T)、腋温(T,临床标准)和直肠温度(T,金标准)。比较了T与T之间的差值以及T与T之间的差值,并根据床型和孕龄对数据进行了分析。T、T和T的平均温度分别为37.16(标准差0.36)℃、36.61(标准差0.30)℃和36.82(标准差0.30)℃。这些方法所测结果均存在显著差异。T - T的平均值为0.21(标准差0.2)℃,T - T的平均值为 - 0.34(标准差0.37)℃,这表明T比T更接近T(P < 0.0001)。对于婴儿床中的婴儿,T与T的一致性比保温箱中的婴儿更高。在调整床型和体重后,随着孕龄每增加一周,T - T与T - T之间的差异减小0.005℃(P = 0.034)。与金标准T相比,T并不比T更准确。颞动脉体温计对于保温箱中的婴儿不如对婴儿床中的婴儿准确。颞动脉温度的准确性随孕龄增加而提高。

相似文献

1
Temporal Artery Temperature Measurement in the Neonate.新生儿颞动脉温度测量
Am J Perinatol. 2017 Aug;34(10):1026-1031. doi: 10.1055/s-0037-1601440. Epub 2017 Apr 10.
2
Comparison of Axillary and Temporal Artery Thermometry in Preterm Neonates.早产儿腋温与颞动脉测温的比较。
J Obstet Gynecol Neonatal Nurs. 2018 May;47(3):352-361. doi: 10.1016/j.jogn.2018.02.013. Epub 2018 Apr 4.
6
Neonatal infrared axillary thermometry.新生儿红外线腋窝测温法。
Clin Pediatr (Phila). 1999 Jan;38(1):35-40. doi: 10.1177/000992289903800105.

引用本文的文献

1
Temperature Monitoring Devices in Neonates.新生儿体温监测设备
Front Pediatr. 2021 Aug 24;9:732810. doi: 10.3389/fped.2021.732810. eCollection 2021.

本文引用的文献

2
Understanding Bland Altman analysis.理解布兰德-奥特曼分析。
Biochem Med (Zagreb). 2015 Jun 5;25(2):141-51. doi: 10.11613/BM.2015.015. eCollection 2015.
4
Comparison of temporal artery and axillary temperatures in healthy newborns.健康新生儿颞动脉温度与腋温的比较。
J Obstet Gynecol Neonatal Nurs. 2012 May-Jun;41(3):383-8. doi: 10.1111/j.1552-6909.2012.01367.x.
7
ACCIDENTAL PERFORATION OF THE COLON AND RECTUM IN NEWBORN INFANTS.新生儿结肠和直肠的意外穿孔
N Engl J Med. 1965 May 27;272:1097-100. doi: 10.1056/NEJM196505272722104.
10
Infrared tympanic thermometry for neonatal temperature assessment.用于新生儿体温评估的红外鼓膜测温法。
J Obstet Gynecol Neonatal Nurs. 1994 Nov-Dec;23(9):798-804. doi: 10.1111/j.1552-6909.1994.tb01955.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验