Assessment of female breast dose for thoracic cone-beam CT using MOSFET dosimeters.

作者信息

Sun Wenzhao, Wang Bin, Qiu Bo, Liang Jian, Xie Weihao, Deng Xiaowu, Qi Zhenyu

机构信息

Department of Radiation Oncology, Sun Yat-Sen University Cancer Center, Collaborative Innovation Center for Cancer Medicine, State Key Laboratory of Oncology in South China, Guangzhou 510060, China.

出版信息

Oncotarget. 2017 Mar 21;8(12):20179-20186. doi: 10.18632/oncotarget.15555.

Abstract

OBJECTIVE

To assess the breast dose during a routine thoracic cone-beam CT (CBCT) check with the efforts to explore the possible dose reduction strategy.

MATERIALS AND METHODS

Metal oxide semiconductor field-effect transistor (MOSFET) dosimeters were used to measure breast surface doses during a thorax kV CBCT scan in an anthropomorphic phantom. Breast doses for different scanning protocols and breast sizes were compared. Dose reduction was attempted by using partial arc CBCT scan with bowtie filter. The impact of this dose reduction strategy on image registration accuracy was investigated.

RESULTS

The average breast surface doses were 20.02 mGy and 11.65 mGy for thoracic CBCT without filtration and with filtration, respectively. This indicates a dose reduction of 41.8% by use of bowtie filter. It was found 220° partial arc scanning significantly reduced the dose to contralateral breast (44.4% lower than ipsilateral breast), while the image registration accuracy was not compromised.

CONCLUSIONS

Breast dose reduction can be achieved by using ipsilateral 220° partial arc scan with bowtie filter. This strategy also provides sufficient image quality for thorax image registration in daily patient positioning verification.

摘要

目的

评估常规胸部锥形束CT(CBCT)检查期间的乳房剂量,并努力探索可能的剂量降低策略。

材料与方法

在人体模型的胸部千伏CBCT扫描过程中,使用金属氧化物半导体场效应晶体管(MOSFET)剂量计测量乳房表面剂量。比较了不同扫描方案和乳房大小的乳房剂量。尝试通过使用带蝴蝶结滤波器的部分弧形CBCT扫描来降低剂量。研究了这种剂量降低策略对图像配准精度的影响。

结果

未过滤和过滤后的胸部CBCT的平均乳房表面剂量分别为20.02 mGy和11.65 mGy。这表明使用蝴蝶结滤波器可使剂量降低41.8%。发现220°部分弧形扫描可显著降低对侧乳房的剂量(比对侧乳房低44.4%),同时图像配准精度不受影响。

结论

使用带蝴蝶结滤波器的同侧220°部分弧形扫描可实现乳房剂量降低。该策略还为日常患者定位验证中的胸部图像配准提供了足够的图像质量。