Excess Radiation to Newborns Hospitalized in the Intensive Care Unit.

作者信息

Komatsu Cássio Vilela, Silva Cristiane Camargo, Souza Luis Ronan Marquez Ferreira de, Gonçalves Luis Fernando

机构信息

Unidade de Diagnóstico por Imagem, Hospital de Clínicas da Universidade Federal do Triângulo Mineiro - HC/UFTM, Rua Getúlio Guaritá, 130, Nossa Senhora da Abadia, CEP 38025-440, Uberaba, Minas Gerais, Brazil.

出版信息

Radiat Prot Dosimetry. 2017 Dec 1;177(3):331-341. doi: 10.1093/rpd/ncx051.

Abstract

Newborns in intensive care units are generally subjected to a great number of X-rays procedures and to the risk of radiation-induced damage. This study evaluated a possible excess radiation by excess of radiographs, excess of dose per radiograph or under-collimation of the X-ray beam. The mean of X-rays per newborn was 12 (1-65) during a mean hospitalization of 29 d (1-226 d). The mean frequency was 0.8 X-ray exposures per newborn. About 13% of X-rays were performed without a well-defined clinical motivation. The mean entrance surface dose of 72 μGy was higher than in most of comparative studies. Under-collimation caused non-thoracic structures to appear frequently on chest radiographs. This study indicates a possible reduction in X-rays exposures by applying the justification principle for each X-ray procedure, recommends a multiprofessional work in the attempt to dose optimization, and shows need of correctly use of collimation system to avoid irradiation of non-thoracic structures.