早产儿重症监护病房胸部X光检查期间的辐射暴露:体模研究

Radiation exposure during chest X-ray examinations in a premature intensive care unit: phantom studies.

作者信息

Duetting T, Foerste B, Knoch T, Darge K, Troeger J

机构信息

Department of Paediatric Radiology, Ruprecht-Karls University, Heidelberg, Germany.

出版信息

Pediatr Radiol. 1999 Mar;29(3):158-62. doi: 10.1007/s002470050562.

Abstract

BACKGROUND

There are few reports on the radiation dose received by infants, their family and radiographers exposed to scatter radiation in a premature baby intensive care unit.

OBJECTIVE

To evaluate the degree of radiation exposure from diagnostic X-ray examinations with mobile X-ray machines in a premature intensive care unit.

MATERIALS AND METHODS

The radiation exposure of an adjacent newborn, the radiographer and other persons in the room was simulated using phantoms during X-ray examination of the chest using vertical and horizontal beams.

RESULTS

Most of the measured doses were below the registration limit of the measuring apparatus and had to be extrapolated by multiple exposures. Without exception, the maximal doses were significantly lower than the permitted limit for persons not professionally exposed to X-rays.

CONCLUSIONS

Recommendations to avoid unnecessary radiation exposure are given.

摘要

背景

关于早产儿重症监护病房中婴儿、其家人以及受到散射辐射的放射技师所接受的辐射剂量,相关报道较少。

目的

评估早产儿重症监护病房中使用移动X射线机进行诊断性X射线检查时的辐射暴露程度。

材料与方法

在胸部X射线检查中,使用垂直和水平射束,通过模型模拟相邻新生儿、放射技师及房间内其他人的辐射暴露情况。

结果

大多数测量剂量低于测量仪器的记录极限,必须通过多次曝光进行外推。无一例外,最大剂量显著低于未从事专业X射线照射人员的允许极限。

结论

给出了避免不必要辐射暴露的建议。