单焦点非球面散光人工晶状体的体外光学质量:柱镜度数的影响。

In vitro optical quality of monofocal aspheric toric intraocular lenses: effect of cylindrical power.

作者信息

Ferrer-Blasco Teresa, Domínguez-Vicent Alberto, García-Lázaro Santiago, Diez María Amparo, Alfonso José F, Esteve-Taboada José J

机构信息

Optometry Research Group (GIO), Department of Optics and Optometry and Vision Sciences, University of Valencia, C/Dr Moliner, 50, 46100, Burjassot, Spain.

Fernández-Vega Ophthalmological Institute, Oviedo, Spain.

出版信息

Int Ophthalmol. 2018 Jun;38(3):933-941. doi: 10.1007/s10792-017-0537-z. Epub 2017 Apr 25.

Abstract

PURPOSE

To assess the in vitro optical quality of monofocal aspheric toric intraocular lenses (IOLs) as a function of the cylindrical power.

METHODS

The in vitro optical quality of the AcrySof IQ Toric IOLs SN6AT2, SN6AT3, SN6AT4, SN6AT5 and SN6AT6 (Alcon Laboratories Inc., Forth Worth, TX, USA) was assessed with an instrument conceived for measuring Zernike's coefficients at 3.0- and 5.0-mm apertures. As a reference, the aspheric monofocal lens AcrySof IQ Aspheric SN60WF (Alcon Laboratories Inc., Forth Worth, TX, USA) was also measured. The area of visibility and cut-off frequency were used to describe the modulation transfer function (MTF) of each lens; meanwhile, the light in the bucket and the diameter of a circular area centred on the point-spread function (PSF) peak that captures 50% of the light energy were used to describe the PSF of each lens. Finally, an image simulation was computed from the Zernike values with reference purposes.

RESULTS

Small differences were found on the metrics used for describing the MTF and PSF of the lenses at both tested apertures, but these were not statistically significant (p > 0.05). Furthermore, the image simulation showed that these differences would not have clinical relevance at all.

CONCLUSIONS

The optical performance of the AcrySof IQ toric IOLs in terms of MTF and PSF is good and seems to be independent of the cylindrical power and similar to a non-toric aspheric lens.

摘要

目的

评估单焦点非球面散光人工晶状体(IOL)的体外光学质量与柱镜度数的关系。

方法

使用一种用于测量3.0毫米和5.0毫米孔径下泽尼克系数的仪器,评估美国德克萨斯州沃思堡爱尔康实验室公司生产的AcrySof IQ Toric IOLs SN6AT2、SN6AT3、SN6AT4、SN6AT5和SN6AT6的体外光学质量。作为参考,还测量了非球面单焦点晶状体AcrySof IQ Aspheric SN60WF(美国德克萨斯州沃思堡爱尔康实验室公司)。用可见度面积和截止频率描述每个晶状体的调制传递函数(MTF);同时,用桶内光线以及以点扩散函数(PSF)峰值为中心、捕获50%光能的圆形区域直径描述每个晶状体的PSF。最后,根据泽尼克值进行图像模拟以供参考。

结果

在两个测试孔径下,用于描述晶状体MTF和PSF的指标存在微小差异,但差异无统计学意义(p>0.05)。此外,图像模拟显示这些差异根本没有临床相关性。

结论

AcrySof IQ散光IOL在MTF和PSF方面的光学性能良好,似乎与柱镜度数无关,且与非散光非球面晶状体相似。