Kidoh Masafumi, Nakaura Takeshi, Funama Yoshinori, Shimonobo Toshiaki, Shirasaka Takashi, Hatemura Masahiro, Utsunomiya Daisuke, Oda Seitaro, Yuki Hideaki, Namimoto Tomohiro, Higaki Toru, Awai Kazuo, Yamashita Yasuyuki
From the Departments of *Diagnostic Radiology and †Medical Physics, Kumamoto University, Kumamoto; and ‡Department of Diagnostic Radiology, Institute of Biomedical and Health Sciences, Hiroshima University, Hiroshima, Japan.
J Comput Assist Tomogr. 2017 May/Jun;41(3):349-353. doi: 10.1097/RCT.0000000000000541.
The aim of this study was to evaluate the effect of cardiac output (CO) on aortic peak enhancement using protocols with different contrast material (CM) injection durations.
We used a flow phantom that simulated the human circulatory system. Contrast material was injected at a rate of 4.0 mL/s for a period of 2.5, 5, 10, 15, or 20 seconds for a CO of 2.8, 4.2, and 5.6 L/min. Single-level serial computed tomography scans of the simulated aorta were acquired after the start of CM delivery, and aortic peak enhancement was recorded under the different injection protocols.
Under a long injection duration protocol (20 seconds), a decrease in CO increased aortic peak enhancement proportionally (CO of 2.8 L/min, 420 Hounsfield units [HU]; CO of 4.2 L/min, 365 HU; CO of 5.6 L/min, 291 HU). However, this effect was decreased under shorter injection duration protocols (5, 10, and 15 seconds); under the shortest (2.5-second) injection duration protocol, a decrease in CO resulted in a decrease in aortic peak enhancement (CO of 2.8 L/min, 36 HU; CO of 4.2 L/min, 51 HU; CO of 5.6 L/min, 55 HU).
The magnitude of the effect of CO on aortic peak enhancement depends on the CM injection duration.
本研究旨在使用不同对比剂(CM)注射时长的方案,评估心输出量(CO)对主动脉峰值增强的影响。
我们使用了一个模拟人体循环系统的血流模型。对于2.8、4.2和5.6L/min的心输出量,以4.0mL/s的速率注射对比剂2.5、5、10、15或20秒。在开始注射对比剂后,对模拟主动脉进行单层连续计算机断层扫描,并记录不同注射方案下的主动脉峰值增强情况。
在长注射时长方案(20秒)下,心输出量降低会成比例地增加主动脉峰值增强(心输出量为2.8L/min时,为420亨氏单位[HU];心输出量为4.2L/min时,为365HU;心输出量为5.6L/min时,为291HU)。然而,在较短注射时长方案(5、10和15秒)下,这种影响会减弱;在最短(2.5秒)注射时长方案下,心输出量降低会导致主动脉峰值增强降低(心输出量为2.8L/min时,为36HU;心输出量为4.2L/min时,为51HU;心输出量为5.6L/min时,为55HU)。
心输出量对主动脉峰值增强的影响程度取决于对比剂注射时长。