Sarfehnia A, Seuntjens J
McGill University, Medical Physics Unit, Montreal, QC.
Med Phys. 2008 Jul;35(7Part3):3416. doi: 10.1118/1.2965993.
Parameters influencing the accuracy of absorbed dose measurements for HDR 192Ir brachytherapy using water calorimetry were investigated with the goal to develop a novel primary absorbed dose to water standard. To provide greater stability, flexibility, and accuracy in the source-detector distance d positioning and measurement, a new spring-loaded catheter holder composed of two concentric cylindrical sleeves with multiple orthogonal adjusting screws was developed. The absorbed dose from Nucletron microSelectron-HDR 192Ir brachytherapy sources with air kerma strengths ranging between 21000-38000 U was studied. d is optimized so as to balance signal-to-noise ratio (decreasing with increasing d ) and temperature drift effects resulting from source self-heating. The irradiation times were adjusted to yield a minimum 1 Gy of dose at the measurement point. Successful measurements at d ranging between 25-50 mm were performed. COMSOL MULTIPHYSICS™ software was used to determine the heat loss correction due to conduction defined as the ratio between temperature rise at a point under ideal conditions to realistic conditions (i.e., no conduction). An agreement of better than 6.5% was observed between TG-43 calculated and calorimetrically measured absorbed dose rates. The effects of convection where calculated to be negligible as the glass vessel provides a convective barrier significantly decoupling the water velocity in the interior and exterior of the vessel (water velocities were 1-2 orders of magnitude different). Our work paves the way to successful primary absorbed dose determination for radioactive sources using calorimetric techniques.
研究了影响使用水热法进行高剂量率192铱近距离放射治疗吸收剂量测量准确性的参数,目的是开发一种新的水吸收剂量初级标准。为了在源 - 探测器距离d的定位和测量中提供更高的稳定性、灵活性和准确性,开发了一种新型弹簧加载导管固定器,它由两个带有多个正交调节螺钉的同心圆柱套筒组成。研究了空气比释动能强度在21000 - 38000 U之间的Nucletron microSelectron - HDR 192铱近距离放射治疗源的吸收剂量。对d进行了优化,以平衡信噪比(随d增加而降低)和源自热导致的温度漂移效应。调整照射时间,使测量点处的剂量至少为1 Gy。在25 - 50 mm的d范围内成功进行了测量。使用COMSOL MULTIPHYSICS™软件确定由于传导引起的热损失校正,传导定义为理想条件下一点的温度上升与实际条件下(即无传导)的温度上升之比。在TG - 43计算的和量热法测量的吸收剂量率之间观察到优于6.5%的一致性。计算得出对流的影响可忽略不计,因为玻璃容器提供了一个对流屏障,使容器内部和外部的水流速度显著解耦(水流速度相差1 - 2个数量级)。我们的工作为使用量热技术成功确定放射源的初级吸收剂量铺平了道路。