Liu D, Reniers B, Poon E, Bazalova M, Rusch T, Verhaegen F
McGill University.
Xoft Inc.
Med Phys. 2008 Jul;35(7Part3):3416. doi: 10.1118/1.2965996.
The Axxent developed by Xoft Inc. is a miniature x-ray tube capable of generating a 50 kVp x-ray spectrum with dose-rates suitable for HDR applications. Results of spectral measurements compared with Geant4 Monte Carlo simulations have been published. This study is a continuation of previous work with shifting emphasis towards dosimetric characterization of the miniature x-ray tube. Dose distributions using EBT Gafchromic films agree to within 10 % of Geant4 results. In addition, TG-43 parameters can be calculated. However, consideration should be given to the biological effectiveness of the spectrum at different depths. Spectral measurements show significant beam hardening with 1 HVL increasing from 0.55 to 1.20 mm Al after 11.50 mm of water filtration. This effect may be attributed to the significant loss of low energy characteristic photons. Furthermore, the degree of beam hardening is dependent of the material, with 1 HVLs of 1.20 and 1.03 mm Al after 11.50 mm of water and Lucite respectively. The biological effect is quantified by calculating the number of single and double strand breaks. The number of strand breaks for the 50 kVp x-ray spectrum is similar to that of I-125 radiation.
Xoft公司研发的Axxent是一种微型X射线管,能够产生50 kVp的X射线能谱,其剂量率适用于高剂量率近距离放射治疗(HDR)应用。已发表了将光谱测量结果与Geant4蒙特卡罗模拟结果进行比较的相关内容。本研究是先前工作的延续,重点转向微型X射线管的剂量学特性研究。使用EBT变色膜得到的剂量分布与Geant4结果的偏差在10%以内。此外,可以计算出TG-43参数。然而,应考虑不同深度处能谱的生物学效应。光谱测量显示,经过11.50 mm水过滤后,半值层(HVL)从0.55 mm铝增加到1.20 mm铝,出现了显著的射线硬化现象。这种效应可能归因于低能特征光子的大量损失。此外,射线硬化程度取决于材料,经过11.50 mm水和有机玻璃过滤后,半值层分别为1.20 mm铝和1.03 mm铝。通过计算单链和双链断裂的数量来量化生物学效应。50 kVp X射线能谱的链断裂数量与碘-125辐射的相似。