Near infrared (NIR) spectroscopy is increasingly gaining significance in the pharmaceutical industry for quality and in-process control. However, the potential of this method for quantitative quality control in pharmacies has long been neglected and little data is available on its application in analysis of creams and ointments. This study evaluated the applicability of NIR spectrometer with limited wavelength range (1000-1900nm) for quantitative quality control of six different dermatological semi-solid pharmaceutical preparations. Each contained a frequently used active ingredient in a common concentration either in a water-free lipid base or in an aqueous cream matrix. Based on direct NIR transflectance measurements through standardized glass beakers and partial least squares (PLS) multivariate calibration, quantitative models were generated comparing several data pre-processing methods Whereas difficulties were observed for mixtures containing 2% (w/w) metronidazole or 4% (w/w) erythromycin, content determination was possible with sufficient accuracy for salicylic acid (5 % (w/w)) and urea (10% (w/w)) in hydrophilic as well as in lipophilic formulations meeting the limit of a maximum deviation of±5% (relative) from the reference values. Exemplarily, one of the methods was successfully validated according to the EMA Guideline, determining several figures of merit such as specificity, linearity, accuracy, precision and robustness.
近红外(NIR)光谱法在制药行业的质量控制和过程控制中越来越重要。然而,该方法在药房定量质量控制方面的潜力长期以来一直被忽视,关于其在乳膏和软膏分析中的应用数据很少。本研究评估了波长范围有限(1000 - 1900nm)的近红外光谱仪对六种不同皮肤科半固体制剂进行定量质量控制的适用性。每种制剂在无水脂质基质或水性乳膏基质中都含有一种常用活性成分,且浓度相同。基于通过标准化玻璃烧杯进行的直接近红外漫反射测量和偏最小二乘法(PLS)多元校准,生成了比较几种数据预处理方法的定量模型。虽然对于含有2%(w/w)甲硝唑或4%(w/w)红霉素的混合物存在困难,但对于亲水性和疏水性制剂中水杨酸(5%(w/w))和尿素(10%(w/w))的含量测定,其准确度足以满足与参考值最大偏差±5%(相对)的限度要求。例如,其中一种方法根据欧洲药品管理局(EMA)指南成功进行了验证,确定了特异性、线性、准确度、精密度和稳健性等几个品质因数。