Doggui Radhouene, Abdelhafidh Sahli Chaïma, Aissa Wassef Lotfi, Hammami Maroua, Ben Sedrine Maha, Mahjoub Rahma, Zouaoui Khemais, Daboubi Rim, Siala Hajer, Messaoud Taieb, Bibi Amina
SURVEN Laboratory, National Institute of Nutrition and Food Technology, Tunis, Tunisia.
Faculty of Sciences of Bizerte, Jarzouna, Tunisia.
Electrophoresis. 2017 Sep;38(17):2210-2218. doi: 10.1002/elps.201600423. Epub 2017 Jun 12.
HbA is used for monitoring diabetic balance. In this paper we report an assessment of the analytical performances of Capillarys 2 Flex Piercing (C2FP) for HbA measurement using CE (Capillary Electrophoresis). CLSI (Clinical and Laboratory Standard Institute) protocols are used for the evaluation of apparatus performances: precision, linearity, method comparison, trueness and common interferences. HbA CVs average in intra-assay was 1.6% between run imprecision CV ranged from 0.1 to 1.8%. The linearity was demonstrated between 4.7 and 15.0%. The comparison study revealed that Bland Altman plot mean difference was equal to -0.03 (CI 95% (-0.05 to -0.0003)) and Passing-Bablok regression intercept was -0.05, CI95%(-0.13 - -0.05); slope: 1.00, CI95%[1.00-1.01]. A strong correlation (r > 0.99) was proved. No significant effects of hemoglobin variants were seen with CE on HbA measurement. No problem related to sample-to-sample carry over was noted. No interferences of LA and cHb were observed. CE allowed quantification of HbA even at low level of total hemoglobin (40 g/L) in contrast to HPLC. Furthermore, this analyzer offered the opportunity of quantifying the HbA simultaneously with HbA . This evaluation showed that C2FP is a convenient system for the control of diabetes and the detection of hemoglobinopathies.
糖化血红蛋白A1(HbA)用于监测糖尿病病情。在本文中,我们报告了使用毛细管电泳(CE)对Capillarys 2 Flex Piercing(C2FP)进行HbA测量的分析性能评估。临床和实验室标准协会(CLSI)的方案用于评估仪器性能:精密度、线性、方法比较、准确性和常见干扰。批内变异系数(CV)平均值为1.6%,批间不精密度CV范围为0.1%至1.8%。线性范围在4.7%至15.0%之间得到验证。比较研究表明,Bland Altman图的平均差异等于-0.03(95%置信区间(-0.05至-0.0003)),Passing-Bablok回归截距为-0.05,95%置信区间(-0.13至-0.05);斜率:1.00,95%置信区间[1.00至1.01]。证明了存在强相关性(r>0.99)。毛细管电泳未发现血红蛋白变异体对HbA测量有显著影响。未发现与样本间携带污染相关的问题。未观察到乳酸(LA)和碳氧血红蛋白(cHb)的干扰。与高效液相色谱(HPLC)相比,即使在总血红蛋白水平较低(40 g/L)时,毛细管电泳也能对HbA进行定量。此外,该分析仪还提供了同时定量HbA1c和HbA的机会。该评估表明,C2FP是一种用于糖尿病控制和血红蛋白病检测的便捷系统。